Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up
Mengmeng Wang,1,* Christine Carole C Corpuz,1 Megumi Fujiwara,1 Minoru Tomita1,2,*1Shinagawa LASIK Center, Tokyo, Japan; 2Department of Ophthalmology, Wenzhou Medical College, Wenzhou, People’s Republic of China*These authors contributed equally to this workPurpose: To e...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Dove Medical Press
2014-05-01
|
Series: | Clinical Ophthalmology |
Online Access: | http://www.dovepress.com/visual-and-optical-performance-of-diffractive-multifocal-intraocular-l-a16818 |
id |
doaj-42a33c74a0c743f5ad6579699f90b06e |
---|---|
record_format |
Article |
spelling |
doaj-42a33c74a0c743f5ad6579699f90b06e2020-11-24T20:46:41ZengDove Medical PressClinical Ophthalmology1177-54832014-05-012014default91992616818Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-upWang MCorpuz CCCFujiwara MTomita M Mengmeng Wang,1,* Christine Carole C Corpuz,1 Megumi Fujiwara,1 Minoru Tomita1,2,*1Shinagawa LASIK Center, Tokyo, Japan; 2Department of Ophthalmology, Wenzhou Medical College, Wenzhou, People’s Republic of China*These authors contributed equally to this workPurpose: To evaluate and compare the visual acuity outcomes and optical performances of eyes implanted with two diffractive multifocal intraocular lens (IOL) models with either a plate haptic design or a modified-C design.Methods: This retrospective study comprised cataract patients who were implanted with either a plate haptic multifocal IOL model (AcrivaUD Reviol BB MFM 611 [VSY Biotechnology, Amsterdam, the Netherlands], group 1) or a modified-C haptic multifocal IOL model ( AcrivaUD Reviol BB MF 613 [VSY Biotechnology, Amsterdam, the Netherlands], group 2) between June 2012 and May 2013. The 6 month postoperative visual acuity, refraction, defocus curve, contrast sensitivity, and wave-front aberration were evaluated and compared between these eyes, using different IOL models.Results: One hundred fifty-eight eyes of 107 patients were included in this study. Significant improvement in visual acuities and refraction was found in both groups after cataract surgery (P<0.01). The visual acuity and contrast sensitivity were statistically better in group 1 than in group 2 (P<0.01). No statistically significant difference in the corneal higher-order aberrations was found between the two groups (P>0.05). However, the ocular higher-order aberrations in group 2 were significantly greater than in group 1 (P<0.05).Conclusion: At 6 months postoperatively, both AcrivaUD Reviol BB MFM 611 IOL and AcrivaUD Reviol BB MF 613 IOL achieved excellent visual and refractive outcomes. The multifocal IOL model with plate haptic design resulted in better optical performances than that with the modified-C haptic design.Keywords: AcrivaUD, VSY, plate haptic, modified-C haptic, multifocal intraocular lens, cataract surgeryhttp://www.dovepress.com/visual-and-optical-performance-of-diffractive-multifocal-intraocular-l-a16818 |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Wang M Corpuz CCC Fujiwara M Tomita M |
spellingShingle |
Wang M Corpuz CCC Fujiwara M Tomita M Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up Clinical Ophthalmology |
author_facet |
Wang M Corpuz CCC Fujiwara M Tomita M |
author_sort |
Wang M |
title |
Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up |
title_short |
Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up |
title_full |
Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up |
title_fullStr |
Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up |
title_full_unstemmed |
Visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up |
title_sort |
visual and optical performance of diffractive multifocal intraocular lenses with different haptic designs: 6 month follow-up |
publisher |
Dove Medical Press |
series |
Clinical Ophthalmology |
issn |
1177-5483 |
publishDate |
2014-05-01 |
description |
Mengmeng Wang,1,* Christine Carole C Corpuz,1 Megumi Fujiwara,1 Minoru Tomita1,2,*1Shinagawa LASIK Center, Tokyo, Japan; 2Department of Ophthalmology, Wenzhou Medical College, Wenzhou, People’s Republic of China*These authors contributed equally to this workPurpose: To evaluate and compare the visual acuity outcomes and optical performances of eyes implanted with two diffractive multifocal intraocular lens (IOL) models with either a plate haptic design or a modified-C design.Methods: This retrospective study comprised cataract patients who were implanted with either a plate haptic multifocal IOL model (AcrivaUD Reviol BB MFM 611 [VSY Biotechnology, Amsterdam, the Netherlands], group 1) or a modified-C haptic multifocal IOL model ( AcrivaUD Reviol BB MF 613 [VSY Biotechnology, Amsterdam, the Netherlands], group 2) between June 2012 and May 2013. The 6 month postoperative visual acuity, refraction, defocus curve, contrast sensitivity, and wave-front aberration were evaluated and compared between these eyes, using different IOL models.Results: One hundred fifty-eight eyes of 107 patients were included in this study. Significant improvement in visual acuities and refraction was found in both groups after cataract surgery (P<0.01). The visual acuity and contrast sensitivity were statistically better in group 1 than in group 2 (P<0.01). No statistically significant difference in the corneal higher-order aberrations was found between the two groups (P>0.05). However, the ocular higher-order aberrations in group 2 were significantly greater than in group 1 (P<0.05).Conclusion: At 6 months postoperatively, both AcrivaUD Reviol BB MFM 611 IOL and AcrivaUD Reviol BB MF 613 IOL achieved excellent visual and refractive outcomes. The multifocal IOL model with plate haptic design resulted in better optical performances than that with the modified-C haptic design.Keywords: AcrivaUD, VSY, plate haptic, modified-C haptic, multifocal intraocular lens, cataract surgery |
url |
http://www.dovepress.com/visual-and-optical-performance-of-diffractive-multifocal-intraocular-l-a16818 |
work_keys_str_mv |
AT wangm visualandopticalperformanceofdiffractivemultifocalintraocularlenseswithdifferenthapticdesignsnbsp6nbspmonthfollowup AT corpuzccc visualandopticalperformanceofdiffractivemultifocalintraocularlenseswithdifferenthapticdesignsnbsp6nbspmonthfollowup AT fujiwaram visualandopticalperformanceofdiffractivemultifocalintraocularlenseswithdifferenthapticdesignsnbsp6nbspmonthfollowup AT tomitam visualandopticalperformanceofdiffractivemultifocalintraocularlenseswithdifferenthapticdesignsnbsp6nbspmonthfollowup |
_version_ |
1716811781035261952 |