Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates
Abstract Background When approaching a joint replacement procedure, pre-surgical planning is essential to predict an accurate estimation of implant size and position. There are currently two methods to achieve it, analog and digital. The present study aims to demonstrate how the hybrid technique is...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
BMC
2021-07-01
|
Series: | Journal of Orthopaedic Surgery and Research |
Subjects: | |
Online Access: | https://doi.org/10.1186/s13018-021-02584-2 |
id |
doaj-4227d63aa6b8483c825f7c33cb252562 |
---|---|
record_format |
Article |
spelling |
doaj-4227d63aa6b8483c825f7c33cb2525622021-07-04T11:38:12ZengBMCJournal of Orthopaedic Surgery and Research1749-799X2021-07-011611610.1186/s13018-021-02584-2Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetatesLuis Fernando Useche Gómez0Hernando Gaitán-Lee1María Alejandra Duarte2Patrick Dennis Halley3Alejandro Romero Jaramillo4Efraim Leal García5Pontificia Universidad Javeriana, Hospital Universitario San IgnacioPontificia Universidad Javeriana, Hospital Universitario San IgnacioPontificia Universidad Javeriana, Hospital Universitario San IgnacioPontificia Universidad Javeriana, Hospital Universitario San IgnacioPontificia Universidad Javeriana, Hospital Universitario San IgnacioPontificia Universidad Javeriana, Hospital Universitario San IgnacioAbstract Background When approaching a joint replacement procedure, pre-surgical planning is essential to predict an accurate estimation of implant size and position. There are currently two methods to achieve it, analog and digital. The present study aims to demonstrate how the hybrid technique is accurate and precise for pre-surgical planning in a non-cemented total hip replacement. Methods Concordance-type study is used against a gold standard, as well as inter- and intra-observer consistency evaluation of two orthopedic surgeons and two orthopedic surgery residents. Accuracy was calculated with the intra-class correlation coefficient (ICC). Afterwards, the same calculation was done considering a margin of error with one size more and one less. Results Thirty-eight patients were included in the study: 19 women and 19 men. Twenty-two prostheses (57.89%) were right-sided and 16 were left (42.11%). Twelve prostheses (31.57%) were Stryker and 26 Johnson & Johnson (68.43%). Acetabular cup correlation compared with the gold standard was moderate: ICC reported 0.45 (95% CI, 0.15–0.76). When adjusted by ± 1 size, ICC was 0.48 (95% CI, 0.18–0.79). On the other hand, results from the femoral stem reported ICC 0.85 (95% CI, 0.07–0.98). When adjusted by ± 1 size, ICC was 0.86 (95% CI, 0.06–0.99). Conclusions Hybrid templating is a reliable substitute for analog or digital planning. It is quick, inexpensive, accurate, and better results are observed in the femoral component regardless the level of expertise of the evaluator. Level of evidence Grade IVhttps://doi.org/10.1186/s13018-021-02584-2Digital templatingAnalog templatingPreoperative planningTotal hip arthroplastyTotal hip replacementReliability |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Luis Fernando Useche Gómez Hernando Gaitán-Lee María Alejandra Duarte Patrick Dennis Halley Alejandro Romero Jaramillo Efraim Leal García |
spellingShingle |
Luis Fernando Useche Gómez Hernando Gaitán-Lee María Alejandra Duarte Patrick Dennis Halley Alejandro Romero Jaramillo Efraim Leal García Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates Journal of Orthopaedic Surgery and Research Digital templating Analog templating Preoperative planning Total hip arthroplasty Total hip replacement Reliability |
author_facet |
Luis Fernando Useche Gómez Hernando Gaitán-Lee María Alejandra Duarte Patrick Dennis Halley Alejandro Romero Jaramillo Efraim Leal García |
author_sort |
Luis Fernando Useche Gómez |
title |
Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates |
title_short |
Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates |
title_full |
Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates |
title_fullStr |
Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates |
title_full_unstemmed |
Precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates |
title_sort |
precision and accuracy of pre-surgical planning of non-cemented total hip replacement with calibrated digital images and acetates |
publisher |
BMC |
series |
Journal of Orthopaedic Surgery and Research |
issn |
1749-799X |
publishDate |
2021-07-01 |
description |
Abstract Background When approaching a joint replacement procedure, pre-surgical planning is essential to predict an accurate estimation of implant size and position. There are currently two methods to achieve it, analog and digital. The present study aims to demonstrate how the hybrid technique is accurate and precise for pre-surgical planning in a non-cemented total hip replacement. Methods Concordance-type study is used against a gold standard, as well as inter- and intra-observer consistency evaluation of two orthopedic surgeons and two orthopedic surgery residents. Accuracy was calculated with the intra-class correlation coefficient (ICC). Afterwards, the same calculation was done considering a margin of error with one size more and one less. Results Thirty-eight patients were included in the study: 19 women and 19 men. Twenty-two prostheses (57.89%) were right-sided and 16 were left (42.11%). Twelve prostheses (31.57%) were Stryker and 26 Johnson & Johnson (68.43%). Acetabular cup correlation compared with the gold standard was moderate: ICC reported 0.45 (95% CI, 0.15–0.76). When adjusted by ± 1 size, ICC was 0.48 (95% CI, 0.18–0.79). On the other hand, results from the femoral stem reported ICC 0.85 (95% CI, 0.07–0.98). When adjusted by ± 1 size, ICC was 0.86 (95% CI, 0.06–0.99). Conclusions Hybrid templating is a reliable substitute for analog or digital planning. It is quick, inexpensive, accurate, and better results are observed in the femoral component regardless the level of expertise of the evaluator. Level of evidence Grade IV |
topic |
Digital templating Analog templating Preoperative planning Total hip arthroplasty Total hip replacement Reliability |
url |
https://doi.org/10.1186/s13018-021-02584-2 |
work_keys_str_mv |
AT luisfernandousechegomez precisionandaccuracyofpresurgicalplanningofnoncementedtotalhipreplacementwithcalibrateddigitalimagesandacetates AT hernandogaitanlee precisionandaccuracyofpresurgicalplanningofnoncementedtotalhipreplacementwithcalibrateddigitalimagesandacetates AT mariaalejandraduarte precisionandaccuracyofpresurgicalplanningofnoncementedtotalhipreplacementwithcalibrateddigitalimagesandacetates AT patrickdennishalley precisionandaccuracyofpresurgicalplanningofnoncementedtotalhipreplacementwithcalibrateddigitalimagesandacetates AT alejandroromerojaramillo precisionandaccuracyofpresurgicalplanningofnoncementedtotalhipreplacementwithcalibrateddigitalimagesandacetates AT efraimlealgarcia precisionandaccuracyofpresurgicalplanningofnoncementedtotalhipreplacementwithcalibrateddigitalimagesandacetates |
_version_ |
1721320064984023040 |