Effect of various physical parameters on surface and build-up dose for 15-MV X-rays
The purpose of this study was to find out the effect of various physical parameters on the skin and build-up doses of 15-MV photon beams. The effects of field dimensions, acrylic shadow tray, focus to-skin distance (FSD) on surface and buildup dose were determined for open, motorized 60° we...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2010-01-01
|
Series: | Journal of Medical Physics |
Subjects: | |
Online Access: | http://www.jmp.org.in/article.asp?issn=0971-6203;year=2010;volume=35;issue=4;spage=202;epage=206;aulast=Yadav |
id |
doaj-3fe63aeb338b4ff4b0b131d62a170f3a |
---|---|
record_format |
Article |
spelling |
doaj-3fe63aeb338b4ff4b0b131d62a170f3a2020-11-24T22:52:38ZengWolters Kluwer Medknow PublicationsJournal of Medical Physics0971-62031998-39132010-01-01354202206Effect of various physical parameters on surface and build-up dose for 15-MV X-raysYadav GirigeshYadav RKumar AlokThe purpose of this study was to find out the effect of various physical parameters on the skin and build-up doses of 15-MV photon beams. The effects of field dimensions, acrylic shadow tray, focus to-skin distance (FSD) on surface and buildup dose were determined for open, motorized 60° wedge (MW) and blocked fields. A ′Markus′ plane parallel plate chamber was used for these measurements in an Elekta (6-15MV) linear accelerator. The surface dose for MW fields was lower than the dose for an open field, but the trend reversed for large fields and higher degree wedges. With the use of an acrylic shadow tray, the surface dose increased for all field sizes, but the increase was dominant for large fields. The surface dose for blocked fields was lower than the dose for open fields. The percentage depth dose of 10 Χ 10 cm <sup>2</sup> field at surface (PDD<sub>0</sub>) for open beam were 13.89%, 11.71%, and 10.74% at 80 cm, 100 cm, and 120 cm FSD, respectively. The blocking tray increased PDD<sub>0</sub> of 10 Χ 10 cm <sup>2</sup> field to 26.29%, 14.01%, and 11.53%, while the motorized 60° wedge decreased PDD<sub>0</sub> to 11.32%, 9.7%, and 8.9 % at these FSDs. The maximum PDD difference seen at surface (i.e., skin) for 5 Χ 5 cm <sup>2</sup> , 15 Χ 15 cm <sup>2</sup> , and 30 Χ 30 cm <sup>2</sup> are 0.5%, 4.6%, and 5.6% for open field and 0.9%, 4.7%, and 7.2% for motorized 60΀ wedge field, when FSDs varied from 80 cm to 120 cm. The maximum PDD difference seen at surface for 5 Χ 5 cm <sup>2</sup> , 15 Χ 15 cm <sup>2</sup> , and 30 Χ 30 cm <sup>2</sup> fields are 5.6%, 22.8%, and 29.6%, respectively, for a 1.0-cm perspex-blocking tray as the FSD is changed. The maximum PDD difference was seen at the surface (i.e., skin) and this decreased with increasing depth.http://www.jmp.org.in/article.asp?issn=0971-6203;year=2010;volume=35;issue=4;spage=202;epage=206;aulast=YadavFocus-to-skin distancemotorized 60° wedgepercentage depth dose at surfacePerspex block traysurface dose |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Yadav Girigesh Yadav R Kumar Alok |
spellingShingle |
Yadav Girigesh Yadav R Kumar Alok Effect of various physical parameters on surface and build-up dose for 15-MV X-rays Journal of Medical Physics Focus-to-skin distance motorized 60° wedge percentage depth dose at surface Perspex block tray surface dose |
author_facet |
Yadav Girigesh Yadav R Kumar Alok |
author_sort |
Yadav Girigesh |
title |
Effect of various physical parameters on surface and build-up dose for 15-MV X-rays |
title_short |
Effect of various physical parameters on surface and build-up dose for 15-MV X-rays |
title_full |
Effect of various physical parameters on surface and build-up dose for 15-MV X-rays |
title_fullStr |
Effect of various physical parameters on surface and build-up dose for 15-MV X-rays |
title_full_unstemmed |
Effect of various physical parameters on surface and build-up dose for 15-MV X-rays |
title_sort |
effect of various physical parameters on surface and build-up dose for 15-mv x-rays |
publisher |
Wolters Kluwer Medknow Publications |
series |
Journal of Medical Physics |
issn |
0971-6203 1998-3913 |
publishDate |
2010-01-01 |
description |
The purpose of this study was to find out the effect of various physical parameters on the skin and build-up doses of 15-MV photon beams. The effects of field dimensions, acrylic shadow tray, focus to-skin distance (FSD) on surface and buildup dose were determined for open, motorized 60° wedge (MW) and blocked fields. A ′Markus′ plane parallel plate chamber was used for these measurements in an Elekta (6-15MV) linear accelerator. The surface dose for MW fields was lower than the dose for an open field, but the trend reversed for large fields and higher degree wedges. With the use of an acrylic shadow tray, the surface dose increased for all field sizes, but the increase was dominant for large fields. The surface dose for blocked fields was lower than the dose for open fields. The percentage depth dose of 10 Χ 10 cm <sup>2</sup> field at surface (PDD<sub>0</sub>) for open beam were 13.89%, 11.71%, and 10.74% at 80 cm, 100 cm, and 120 cm FSD, respectively. The blocking tray increased PDD<sub>0</sub> of 10 Χ 10 cm <sup>2</sup> field to 26.29%, 14.01%, and 11.53%, while the motorized 60° wedge decreased PDD<sub>0</sub> to 11.32%, 9.7%, and 8.9 % at these FSDs. The maximum PDD difference seen at surface (i.e., skin) for 5 Χ 5 cm <sup>2</sup> , 15 Χ 15 cm <sup>2</sup> , and 30 Χ 30 cm <sup>2</sup> are 0.5%, 4.6%, and 5.6% for open field and 0.9%, 4.7%, and 7.2% for motorized 60΀ wedge field, when FSDs varied from 80 cm to 120 cm. The maximum PDD difference seen at surface for 5 Χ 5 cm <sup>2</sup> , 15 Χ 15 cm <sup>2</sup> , and 30 Χ 30 cm <sup>2</sup> fields are 5.6%, 22.8%, and 29.6%, respectively, for a 1.0-cm perspex-blocking tray as the FSD is changed. The maximum PDD difference was seen at the surface (i.e., skin) and this decreased with increasing depth. |
topic |
Focus-to-skin distance motorized 60° wedge percentage depth dose at surface Perspex block tray surface dose |
url |
http://www.jmp.org.in/article.asp?issn=0971-6203;year=2010;volume=35;issue=4;spage=202;epage=206;aulast=Yadav |
work_keys_str_mv |
AT yadavgirigesh effectofvariousphysicalparametersonsurfaceandbuildupdosefor15mvxrays AT yadavr effectofvariousphysicalparametersonsurfaceandbuildupdosefor15mvxrays AT kumaralok effectofvariousphysicalparametersonsurfaceandbuildupdosefor15mvxrays |
_version_ |
1725665310045896704 |