Summary: | Abstract Purpose Ultrashort echo time (UTE) sequences play a key role in imaging and quantifying short T2 species. However, almost all of the relevant studies was conducted at relatively high fields. The purpose of this work was to further explore the feasibility of UTE imaging and T2* measurement for short T2 species at low fields. Methods A 2D UTE sequence with an echo time (TE) of 0.37 ms was developed on a 0.35 T permanent magnet scanner. This sequence acquires multiecho images to fit the monoexponential signal decay model for quantitative T2* calculations. In the phantom experiments, MnCl2 solutions with different T2* values were used to assess the curve fitting model in low fields. In the in vivo experiments, T2* measurements were performed on the Achilles tendon of five normal volunteers. Results The phantom studies showed a significant linear relationship between the MnCl2 solution concentration and R2* (1/T2*) values, which indicated the stability and accuracy of the T2* quantification model. The in vivo studies demonstrated that mean T2* value of Achilles tendon is 1.83 ± 0.21 ms, and the mean coefficient of determination (R-squared) was 0.996. Conclusions Both phantom and in vivo experiments showed that UTE imaging and quantification for short T2 components were feasible at low field 0.35 T scanner. This pilot study presents preliminary conclusions for future work.
|