A case of ventricular fibrillation not detected by an automated external defibrillator
We report a case of ventricular fibrillation (VF) which an automated external defibrillator (AED) could not detect. A 13-year-old girl collapsed just after playing basketball. Cardio pulmonary resuscitation (CPR) was started immediately and 5 minutes later an AED was applied by a rescue team. The mo...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wiley
2008-01-01
|
Series: | Journal of Arrhythmia |
Subjects: | |
Online Access: | http://www.sciencedirect.com/science/article/pii/S1880427608800257 |
id |
doaj-3c57ab6a86524aba88d7b608dfa22b70 |
---|---|
record_format |
Article |
spelling |
doaj-3c57ab6a86524aba88d7b608dfa22b702020-11-24T22:38:55ZengWileyJournal of Arrhythmia1880-42762008-01-0124316616910.1016/S1880-4276(08)80025-7A case of ventricular fibrillation not detected by an automated external defibrillatorShigetoshi Sakabe, MDAtsunobu Kasai, MDYuichi Sato, MDTakashi Omura, MDTakashi Yamanaka, MDMorimichi Setsuda, MDWe report a case of ventricular fibrillation (VF) which an automated external defibrillator (AED) could not detect. A 13-year-old girl collapsed just after playing basketball. Cardio pulmonary resuscitation (CPR) was started immediately and 5 minutes later an AED was applied by a rescue team. The monitor of the AED displayed typical VF, but the AED did not detect it as VF. The VF was not detected during 2 more attempts. Detection occurred on the fourth attempt, and counter-shock was successfully delivered, but the process took an extra 9 minutes. After the event, the girl was diagnosed with a latent type of prolonged QT syndrome. We analyzed the reason why the VF was not detected by the AED and found that the ECG detected by the AED fell outside the device’s parameters for ventricular tachycardia (VT) or VF. We emphasize that the AED is an excellent device, but we should also be aware of its limitations.http://www.sciencedirect.com/science/article/pii/S1880427608800257Automated external defibrillationVentricular fibrillationIdentification |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Shigetoshi Sakabe, MD Atsunobu Kasai, MD Yuichi Sato, MD Takashi Omura, MD Takashi Yamanaka, MD Morimichi Setsuda, MD |
spellingShingle |
Shigetoshi Sakabe, MD Atsunobu Kasai, MD Yuichi Sato, MD Takashi Omura, MD Takashi Yamanaka, MD Morimichi Setsuda, MD A case of ventricular fibrillation not detected by an automated external defibrillator Journal of Arrhythmia Automated external defibrillation Ventricular fibrillation Identification |
author_facet |
Shigetoshi Sakabe, MD Atsunobu Kasai, MD Yuichi Sato, MD Takashi Omura, MD Takashi Yamanaka, MD Morimichi Setsuda, MD |
author_sort |
Shigetoshi Sakabe, MD |
title |
A case of ventricular fibrillation not detected by an automated external defibrillator |
title_short |
A case of ventricular fibrillation not detected by an automated external defibrillator |
title_full |
A case of ventricular fibrillation not detected by an automated external defibrillator |
title_fullStr |
A case of ventricular fibrillation not detected by an automated external defibrillator |
title_full_unstemmed |
A case of ventricular fibrillation not detected by an automated external defibrillator |
title_sort |
case of ventricular fibrillation not detected by an automated external defibrillator |
publisher |
Wiley |
series |
Journal of Arrhythmia |
issn |
1880-4276 |
publishDate |
2008-01-01 |
description |
We report a case of ventricular fibrillation (VF) which an automated external defibrillator (AED) could not detect. A 13-year-old girl collapsed just after playing basketball. Cardio pulmonary resuscitation (CPR) was started immediately and 5 minutes later an AED was applied by a rescue team. The monitor of the AED displayed typical VF, but the AED did not detect it as VF. The VF was not detected during 2 more attempts. Detection occurred on the fourth attempt, and counter-shock was successfully delivered, but the process took an extra 9 minutes. After the event, the girl was diagnosed with a latent type of prolonged QT syndrome. We analyzed the reason why the VF was not detected by the AED and found that the ECG detected by the AED fell outside the device’s parameters for ventricular tachycardia (VT) or VF. We emphasize that the AED is an excellent device, but we should also be aware of its limitations. |
topic |
Automated external defibrillation Ventricular fibrillation Identification |
url |
http://www.sciencedirect.com/science/article/pii/S1880427608800257 |
work_keys_str_mv |
AT shigetoshisakabemd acaseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT atsunobukasaimd acaseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT yuichisatomd acaseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT takashiomuramd acaseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT takashiyamanakamd acaseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT morimichisetsudamd acaseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT shigetoshisakabemd caseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT atsunobukasaimd caseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT yuichisatomd caseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT takashiomuramd caseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT takashiyamanakamd caseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator AT morimichisetsudamd caseofventricularfibrillationnotdetectedbyanautomatedexternaldefibrillator |
_version_ |
1725711133257498624 |