The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.

<h4>Background</h4>ClearSight is a noninvasive arterial blood pressure monitor, but it remains unknown whether it is affected by the state of perfusion to the fingers. We investigated whether the lower perfusion index (PI) measured with a pulse oximeter, which reflects finger perfusion,...

Full description

Bibliographic Details
Main Authors: Masashi Yokose, Takahiro Mihara, Masahiro Takaya, Takumi Yamamoto, Yusuke Saigusa, Shunsuke Takaki, Takahisa Goto
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2019-01-01
Series:PLoS ONE
Online Access:https://doi.org/10.1371/journal.pone.0219511
id doaj-2f514fcda33048538e7a87ef54723253
record_format Article
spelling doaj-2f514fcda33048538e7a87ef547232532021-03-04T10:28:16ZengPublic Library of Science (PLoS)PLoS ONE1932-62032019-01-01147e021951110.1371/journal.pone.0219511The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.Masashi YokoseTakahiro MiharaMasahiro TakayaTakumi YamamotoYusuke SaigusaShunsuke TakakiTakahisa Goto<h4>Background</h4>ClearSight is a noninvasive arterial blood pressure monitor, but it remains unknown whether it is affected by the state of perfusion to the fingers. We investigated whether the lower perfusion index (PI) measured with a pulse oximeter, which reflects finger perfusion, would affect the agreement between arterial pressures measured with ClearSight versus those obtained with an arterial catheter.<h4>Methods</h4>Paired arterial pressure data (ClearSight and arterial catheter-based pressures) and PI values were prospectively obtained from 30 patients undergoing major abdominal surgery. The primary outcome was standard deviation (SD) of the bias (precision) of blood pressure between ClearSight and arterial catheter. The ratio of the adjusted SD of the bias between PI≤1 and PI>1 was calculated using the linear mixed-effects model. The secondary outcomes were the bias and the limits of agreement (LOA) between the two devices (repeated measures Bland-Altman analysis).<h4>Results</h4>We analyzed 6312 paired data points. The adjusted SD of bias in PI ≤1 compared with those in PI >1 was 1.4-fold (95% confidence interval: 1.3- to 1.4-fold) for systolic arterial pressure, 1.5-fold (95% confidence interval: 1.3- to 1.6-fold) for diastolic arterial pressure, and 1.3-fold (95% confidence interval: 1.2- to 1.5-fold) for mean arterial pressure. The bias (LOA) were as follows: systolic arterial pressure in the PI ≤1 and PI >1 groups, -3.5 (-35.4 to 28.4) mmHg and 2.2 (-19.9 to 24.3) mmHg, respectively; diastolic arterial pressure in the PI ≤1 and PI >1 groups, 13.1 (-5.1 to 31.3) mmHg and 9.0, (-2.6 to 20.6) mmHg, respectively; and mean arterial pressure in the PI ≤1 and PI >1 groups, 8.7 (-11.3 to 28.7) mmHg and 7.6 (-6.2 to 21.3) mmHg, respectively.<h4>Conclusions</h4>PI ≤1 was associated with a large SD of the bias between the devices. The PI value could be a real-time indicator of ClearSight precision.https://doi.org/10.1371/journal.pone.0219511
collection DOAJ
language English
format Article
sources DOAJ
author Masashi Yokose
Takahiro Mihara
Masahiro Takaya
Takumi Yamamoto
Yusuke Saigusa
Shunsuke Takaki
Takahisa Goto
spellingShingle Masashi Yokose
Takahiro Mihara
Masahiro Takaya
Takumi Yamamoto
Yusuke Saigusa
Shunsuke Takaki
Takahisa Goto
The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.
PLoS ONE
author_facet Masashi Yokose
Takahiro Mihara
Masahiro Takaya
Takumi Yamamoto
Yusuke Saigusa
Shunsuke Takaki
Takahisa Goto
author_sort Masashi Yokose
title The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.
title_short The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.
title_full The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.
title_fullStr The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.
title_full_unstemmed The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.
title_sort perfusion index measured by the pulse oximeter affects the agreement between clearsight and the arterial catheter-based blood pressures: a prospective observational study.
publisher Public Library of Science (PLoS)
series PLoS ONE
issn 1932-6203
publishDate 2019-01-01
description <h4>Background</h4>ClearSight is a noninvasive arterial blood pressure monitor, but it remains unknown whether it is affected by the state of perfusion to the fingers. We investigated whether the lower perfusion index (PI) measured with a pulse oximeter, which reflects finger perfusion, would affect the agreement between arterial pressures measured with ClearSight versus those obtained with an arterial catheter.<h4>Methods</h4>Paired arterial pressure data (ClearSight and arterial catheter-based pressures) and PI values were prospectively obtained from 30 patients undergoing major abdominal surgery. The primary outcome was standard deviation (SD) of the bias (precision) of blood pressure between ClearSight and arterial catheter. The ratio of the adjusted SD of the bias between PI≤1 and PI>1 was calculated using the linear mixed-effects model. The secondary outcomes were the bias and the limits of agreement (LOA) between the two devices (repeated measures Bland-Altman analysis).<h4>Results</h4>We analyzed 6312 paired data points. The adjusted SD of bias in PI ≤1 compared with those in PI >1 was 1.4-fold (95% confidence interval: 1.3- to 1.4-fold) for systolic arterial pressure, 1.5-fold (95% confidence interval: 1.3- to 1.6-fold) for diastolic arterial pressure, and 1.3-fold (95% confidence interval: 1.2- to 1.5-fold) for mean arterial pressure. The bias (LOA) were as follows: systolic arterial pressure in the PI ≤1 and PI >1 groups, -3.5 (-35.4 to 28.4) mmHg and 2.2 (-19.9 to 24.3) mmHg, respectively; diastolic arterial pressure in the PI ≤1 and PI >1 groups, 13.1 (-5.1 to 31.3) mmHg and 9.0, (-2.6 to 20.6) mmHg, respectively; and mean arterial pressure in the PI ≤1 and PI >1 groups, 8.7 (-11.3 to 28.7) mmHg and 7.6 (-6.2 to 21.3) mmHg, respectively.<h4>Conclusions</h4>PI ≤1 was associated with a large SD of the bias between the devices. The PI value could be a real-time indicator of ClearSight precision.
url https://doi.org/10.1371/journal.pone.0219511
work_keys_str_mv AT masashiyokose theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT takahiromihara theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT masahirotakaya theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT takumiyamamoto theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT yusukesaigusa theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT shunsuketakaki theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT takahisagoto theperfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT masashiyokose perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT takahiromihara perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT masahirotakaya perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT takumiyamamoto perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT yusukesaigusa perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT shunsuketakaki perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
AT takahisagoto perfusionindexmeasuredbythepulseoximeteraffectstheagreementbetweenclearsightandthearterialcatheterbasedbloodpressuresaprospectiveobservationalstudy
_version_ 1714805806538424320