Complex water effluents pose a toxicity risk to biological wastewater treatments and environmental discharge. Dynamic sampling and risk assessment of effluents would mitigate downstream hazards, but few methods are available to assess microorganism toxicity on the minute timescale. To rapidly evalua...

Full description

Bibliographic Details
Main Authors: Raphael Ortiz, Jian Lin Chen, David C. Stuckey, Terry W.J. Steele
Format: Article
Language:English
Published: Elsevier 2019-03-01
Series:Micro and Nano Engineering
Online Access:http://www.sciencedirect.com/science/article/pii/S2590007219300061
id doaj-2ef31561ce224065bd94590f8d6d3b6f
record_format Article
spelling doaj-2ef31561ce224065bd94590f8d6d3b6f2020-11-25T01:18:01ZengElsevierMicro and Nano Engineering2590-00722019-03-01292103Rapid serial diluting biomicrofluidic provides EC50 in minutesRaphael Ortiz0Jian Lin Chen1David C. Stuckey2Terry W.J. Steele3School of Materials Science & Engineering, College of Engineering, Nanyang Technological University, 50 Nanyang Avenue, 639798, Singapore; Advanced Environmental Biotechnology Centre, Nanyang Environment & Water Research Institute, Nanyang Technological University, 637141, SingaporeAdvanced Environmental Biotechnology Centre, Nanyang Environment & Water Research Institute, Nanyang Technological University, 637141, Singapore; Department of Chemistry, City University of Hong Kong, Hong Kong, SAR, PR ChinaAdvanced Environmental Biotechnology Centre, Nanyang Environment & Water Research Institute, Nanyang Technological University, 637141, Singapore; Chemical Engineering Department, Imperial College London, London SW7 2AZ, UKSchool of Materials Science & Engineering, College of Engineering, Nanyang Technological University, 50 Nanyang Avenue, 639798, Singapore; Corresponding author.Complex water effluents pose a toxicity risk to biological wastewater treatments and environmental discharge. Dynamic sampling and risk assessment of effluents would mitigate downstream hazards, but few methods are available to assess microorganism toxicity on the minute timescale. To rapidly evaluate unknown aqueous effluents, a segmented-flow microfluidic device is refined for real-time gross toxicity detection and quantification. A microfluidic chip integrating a magnetic stirrer for serial dilution is demonstrated to rapidly determine EC50 values of known and unknown toxicants, where 100 nL samples are suspended into a continuous oil phase with no interfering surfactants. A five logarithmic dilution sequence is evaluated in <5 min, enabling feedback for dynamic processes e.g. dilution, ion exchange, or absorbant treatments. The resazurin-based assay is refined from traditional laboratory procedures to reduce sample volume and response time, with the advantages of short toxicant exposure (< 30 s), high density bacteria (10^7 cell/mL), and continuous mixing in an oxygen-free medium. Simple scaling of the stirred chamber volume from 1 to 30 μL adjusts the number of samples from 100 to 2500 dilution droplets, respectively. A detailed analysis of the resazurin kinetics suggests the presence of two or more enzymes with distinct Michaelis constants. Enzyme kinetics and the resazurin reduction rate is dependent on growth phase and EC50 inhibition by mercury > zinc > copper > nickel cations. Resazurin concentration of 10 μM is optimal for Enterococcus faecalis at 0.1 optical density. The adaptable method is transferable to other microorganisms, such as common baker's yeast, Saccharomyces cerevisiae, where the resazurin reduction rate is 30% of Enterococcus faecalis (4 nM/s vs. 13 nM/s per 100 nL droplet). Zinc and nickel cations are observed to increase the base resazurin reduction rate of baker's yeast by 25%, whereas copper is found to be more cytotoxic than mercury cations. Keywords: Microfluidic, Droplet, EC50, Heavy metals, Enzyme kinetics, Resazurinhttp://www.sciencedirect.com/science/article/pii/S2590007219300061
collection DOAJ
language English
format Article
sources DOAJ
author Raphael Ortiz
Jian Lin Chen
David C. Stuckey
Terry W.J. Steele
spellingShingle Raphael Ortiz
Jian Lin Chen
David C. Stuckey
Terry W.J. Steele
Rapid serial diluting biomicrofluidic provides EC50 in minutes
Micro and Nano Engineering
author_facet Raphael Ortiz
Jian Lin Chen
David C. Stuckey
Terry W.J. Steele
author_sort Raphael Ortiz
title Rapid serial diluting biomicrofluidic provides EC50 in minutes
title_short Rapid serial diluting biomicrofluidic provides EC50 in minutes
title_full Rapid serial diluting biomicrofluidic provides EC50 in minutes
title_fullStr Rapid serial diluting biomicrofluidic provides EC50 in minutes
title_full_unstemmed Rapid serial diluting biomicrofluidic provides EC50 in minutes
title_sort rapid serial diluting biomicrofluidic provides ec50 in minutes
publisher Elsevier
series Micro and Nano Engineering
issn 2590-0072
publishDate 2019-03-01
description Complex water effluents pose a toxicity risk to biological wastewater treatments and environmental discharge. Dynamic sampling and risk assessment of effluents would mitigate downstream hazards, but few methods are available to assess microorganism toxicity on the minute timescale. To rapidly evaluate unknown aqueous effluents, a segmented-flow microfluidic device is refined for real-time gross toxicity detection and quantification. A microfluidic chip integrating a magnetic stirrer for serial dilution is demonstrated to rapidly determine EC50 values of known and unknown toxicants, where 100 nL samples are suspended into a continuous oil phase with no interfering surfactants. A five logarithmic dilution sequence is evaluated in <5 min, enabling feedback for dynamic processes e.g. dilution, ion exchange, or absorbant treatments. The resazurin-based assay is refined from traditional laboratory procedures to reduce sample volume and response time, with the advantages of short toxicant exposure (< 30 s), high density bacteria (10^7 cell/mL), and continuous mixing in an oxygen-free medium. Simple scaling of the stirred chamber volume from 1 to 30 μL adjusts the number of samples from 100 to 2500 dilution droplets, respectively. A detailed analysis of the resazurin kinetics suggests the presence of two or more enzymes with distinct Michaelis constants. Enzyme kinetics and the resazurin reduction rate is dependent on growth phase and EC50 inhibition by mercury > zinc > copper > nickel cations. Resazurin concentration of 10 μM is optimal for Enterococcus faecalis at 0.1 optical density. The adaptable method is transferable to other microorganisms, such as common baker's yeast, Saccharomyces cerevisiae, where the resazurin reduction rate is 30% of Enterococcus faecalis (4 nM/s vs. 13 nM/s per 100 nL droplet). Zinc and nickel cations are observed to increase the base resazurin reduction rate of baker's yeast by 25%, whereas copper is found to be more cytotoxic than mercury cations. Keywords: Microfluidic, Droplet, EC50, Heavy metals, Enzyme kinetics, Resazurin
url http://www.sciencedirect.com/science/article/pii/S2590007219300061
work_keys_str_mv AT raphaelortiz rapidserialdilutingbiomicrofluidicprovidesec50inminutes
AT jianlinchen rapidserialdilutingbiomicrofluidicprovidesec50inminutes
AT davidcstuckey rapidserialdilutingbiomicrofluidicprovidesec50inminutes
AT terrywjsteele rapidserialdilutingbiomicrofluidicprovidesec50inminutes
_version_ 1725144266765762560