Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies

Abstract Objective: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. Methods: The systolic and diastolic durations from three in vivo st...

Full description

Bibliographic Details
Main Authors: Ovandir Bazan, Jayme Pinto Ortiz
Format: Article
Language:English
Published: Sociedade Brasileira de Cirurgia Cardiovascular
Series:Brazilian Journal of Cardiovascular Surgery
Subjects:
Online Access:http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000200171&lng=en&tlng=en
id doaj-274c63cb77434cca9ca673b78f148b05
record_format Article
spelling doaj-274c63cb77434cca9ca673b78f148b052020-11-24T22:39:32ZengSociedade Brasileira de Cirurgia CardiovascularBrazilian Journal of Cardiovascular Surgery1678-974131217117310.5935/1678-9741.20160036S0102-76382016000200171Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo StudiesOvandir BazanJayme Pinto OrtizAbstract Objective: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. Methods: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. Results: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. Conclusion: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000200171&lng=en&tlng=enVentricular Function, LeftHeart valve prosthesisIn Vitro TechniquesHeart Rate
collection DOAJ
language English
format Article
sources DOAJ
author Ovandir Bazan
Jayme Pinto Ortiz
spellingShingle Ovandir Bazan
Jayme Pinto Ortiz
Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
Brazilian Journal of Cardiovascular Surgery
Ventricular Function, Left
Heart valve prosthesis
In Vitro Techniques
Heart Rate
author_facet Ovandir Bazan
Jayme Pinto Ortiz
author_sort Ovandir Bazan
title Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_short Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_full Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_fullStr Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_full_unstemmed Duration of Systole and Diastole for Hydrodynamic Testing of Prosthetic Heart Valves: Comparison Between ISO 5840 Standards and in vivo Studies
title_sort duration of systole and diastole for hydrodynamic testing of prosthetic heart valves: comparison between iso 5840 standards and in vivo studies
publisher Sociedade Brasileira de Cirurgia Cardiovascular
series Brazilian Journal of Cardiovascular Surgery
issn 1678-9741
description Abstract Objective: To complement the ISO 5840 standards concerning the duration of left ventricular systole and diastole as a function of changes in heart rates according to in vivo studies from the physiologic literature review. Methods: The systolic and diastolic durations from three in vivo studies were compared with the durations of systole proposed by the ISO 5840:2010 and ISO 5840-2:2015 for hydrodynamic performance assessment of prosthetic heart valves. Results: Based on the in vivo studies analyzed, the systolic durations proposed by the ISO 5840 standard seemed consistent for 45 and 120 beats per minute (bpm), and showed diverse results for the 70 bpm condition. Conclusion: Information on the realistic validation of the operation of left ventricular models for different heart rates were obtained.
topic Ventricular Function, Left
Heart valve prosthesis
In Vitro Techniques
Heart Rate
url http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0102-76382016000200171&lng=en&tlng=en
work_keys_str_mv AT ovandirbazan durationofsystoleanddiastoleforhydrodynamictestingofprostheticheartvalvescomparisonbetweeniso5840standardsandinvivostudies
AT jaymepintoortiz durationofsystoleanddiastoleforhydrodynamictestingofprostheticheartvalvescomparisonbetweeniso5840standardsandinvivostudies
_version_ 1725708431323561984