Summary: | Endoscopic endonasal surgery (EES) is a minimally invasive technique for removal of pituitary adenomas or cysts at the skull base. This approach can reduce the invasiveness and recovery time compared to traditional open surgery techniques. However, it represents challenges to surgeons because of the constrained workspace imposed by the nasal cavity and the lack of dexterity with conventional surgical instruments. While robotic surgical systems have been previously proposed for EES, issues concerned with proper interface design still remain. In this paper, we present a cooperative, compact, and versatile bimanual human-robot interface aimed to provide intuitive and safe operation in robot-assisted EES. The proposed interface is attached to a robot arm and holds a multi-degree-of-freedom (DOF) articulated forceps. In order to design the required functionalities in EES, we consider a simplified surgical task scenario, with four basic instrument operations such as positioning, insertion, manipulation, and extraction. The proposed cooperative strategy is based on the combination of force based robot control for tool positioning, a virtual remote-center-of-motion (VRCM) during insertion/extraction tasks, and the use of a serial-link interface for precise and simultaneous control of the position and the orientation of the forceps tip. Virtual workspace constraints and motion scaling are added to provide safe and smooth control of our robotic surgical system. We evaluate the performance and usability of our system considering reachability, object manipulability, and surgical dexterity in an anatomically realistic human head phantom compared to the use of conventional surgical instruments. The results demonstrate that the proposed system can improve the precision, smoothness and safety of the forceps operation during an EES.
|