Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring

<p><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Introduction:</strong> Transducer calibration is very important for the accurate measurement of pressures during invasive hemodynamic monitoring of patients requiring close o...

Full description

Bibliographic Details
Main Authors: Alberto Hernández González, Lázaro Edel Sánchez Escalona, Caridad de Dios Soler Morejón, Mirtha López Ramírez, Agustín Mulet Gómez
Format: Article
Language:Spanish
Published: Centro Nacional de Información de Ciencias Médicas. Editorial de Ciencias Médicas (ECIMED) 2017-05-01
Series:Revista Cubana de Anestesiología y Reanimación
Subjects:
Online Access:http://revanestesia.sld.cu/index.php/anestRean/article/view/152
id doaj-2246f8c1e8ba49b68b3f46a5060de9f1
record_format Article
spelling doaj-2246f8c1e8ba49b68b3f46a5060de9f12020-11-29T03:49:44ZspaCentro Nacional de Información de Ciencias Médicas. Editorial de Ciencias Médicas (ECIMED)Revista Cubana de Anestesiología y Reanimación1726-67182017-05-01162151Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoringAlberto Hernández González0Lázaro Edel Sánchez Escalona1Caridad de Dios Soler Morejón2Mirtha López Ramírez3Agustín Mulet Gómez4Hospital Clínico Quirúrgico "Hermanos Ameijeiras". La HabanaHospital Clínico Quirúrgico "Hermanos Ameijeiras". La HabanaHospital Clínico Quirúrgico "Hermanos Ameijeiras". La HabanaHospital Clínico Quirúrgico "Hermanos Ameijeiras". La HabanaHospital Clínico Quirúrgico "Hermanos Ameijeiras". La Habana<p><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Introduction:</strong> Transducer calibration is very important for the accurate measurement of pressures during invasive hemodynamic monitoring of patients requiring close observation in the primary care units and in the operating room. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Objective</strong>: To develop and validate a method of calibration of pressure transducers, which is simple, efficient and low-cost. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Methods: </strong> An alternative manometric method, based on Pascal's law, was developed for the calibration of pressure transducers during invasive hemodynamic monitoring performed at the intensive care unit of the cardiovascular surgery department at Hermanos Ameijeiras Clinical-Surgical Hospital. This method was validated by its comparison with the recommended standard in a total of 215 patients who required invasive hemodynamic monitoring between January and June 2015. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Results</strong>: With the alternative manometric method created, the same result was obtained as with the hydraulic method recommended by the literature. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Conclusions:</strong> The alternative method created is accurate, efficient and cost-effective. </span></p>http://revanestesia.sld.cu/index.php/anestRean/article/view/152transductor de presiónmonitorización hemodinámica invasivacalibración de transductores.
collection DOAJ
language Spanish
format Article
sources DOAJ
author Alberto Hernández González
Lázaro Edel Sánchez Escalona
Caridad de Dios Soler Morejón
Mirtha López Ramírez
Agustín Mulet Gómez
spellingShingle Alberto Hernández González
Lázaro Edel Sánchez Escalona
Caridad de Dios Soler Morejón
Mirtha López Ramírez
Agustín Mulet Gómez
Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
Revista Cubana de Anestesiología y Reanimación
transductor de presión
monitorización hemodinámica invasiva
calibración de transductores.
author_facet Alberto Hernández González
Lázaro Edel Sánchez Escalona
Caridad de Dios Soler Morejón
Mirtha López Ramírez
Agustín Mulet Gómez
author_sort Alberto Hernández González
title Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
title_short Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
title_full Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
title_fullStr Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
title_full_unstemmed Alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
title_sort alternative manometric method of calibration of pressure transducers for invasive hemodynamic monitoring
publisher Centro Nacional de Información de Ciencias Médicas. Editorial de Ciencias Médicas (ECIMED)
series Revista Cubana de Anestesiología y Reanimación
issn 1726-6718
publishDate 2017-05-01
description <p><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Introduction:</strong> Transducer calibration is very important for the accurate measurement of pressures during invasive hemodynamic monitoring of patients requiring close observation in the primary care units and in the operating room. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Objective</strong>: To develop and validate a method of calibration of pressure transducers, which is simple, efficient and low-cost. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Methods: </strong> An alternative manometric method, based on Pascal's law, was developed for the calibration of pressure transducers during invasive hemodynamic monitoring performed at the intensive care unit of the cardiovascular surgery department at Hermanos Ameijeiras Clinical-Surgical Hospital. This method was validated by its comparison with the recommended standard in a total of 215 patients who required invasive hemodynamic monitoring between January and June 2015. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Results</strong>: With the alternative manometric method created, the same result was obtained as with the hydraulic method recommended by the literature. <br /> </span><span style="font-family: Verdana,Arial,Helvetica,sans-serif; font-size: small;"><strong>Conclusions:</strong> The alternative method created is accurate, efficient and cost-effective. </span></p>
topic transductor de presión
monitorización hemodinámica invasiva
calibración de transductores.
url http://revanestesia.sld.cu/index.php/anestRean/article/view/152
work_keys_str_mv AT albertohernandezgonzalez alternativemanometricmethodofcalibrationofpressuretransducersforinvasivehemodynamicmonitoring
AT lazaroedelsanchezescalona alternativemanometricmethodofcalibrationofpressuretransducersforinvasivehemodynamicmonitoring
AT caridaddediossolermorejon alternativemanometricmethodofcalibrationofpressuretransducersforinvasivehemodynamicmonitoring
AT mirthalopezramirez alternativemanometricmethodofcalibrationofpressuretransducersforinvasivehemodynamicmonitoring
AT agustinmuletgomez alternativemanometricmethodofcalibrationofpressuretransducersforinvasivehemodynamicmonitoring
_version_ 1724412503024205824