A Validated RP-HPLC Method for the Estimation of Quetiapine in Bulk and Pharmaceutical Formulations

A new, simple, specific, sensitive, rapid, accurate and precise RP-HPLC method was developed for the estimation of quetiapine in bulk and pharmaceutical formulations. Quetiapine was chromatographed on a reverse phase C18 Waters column (75x4.6 mm I.D., particle size 3.5 μm) in a mobile phase consisti...

Full description

Bibliographic Details
Main Authors: D. Suneetha, A. Lakshmana Rao
Format: Article
Language:English
Published: Hindawi Limited 2010-01-01
Series:E-Journal of Chemistry
Online Access:http://dx.doi.org/10.1155/2010/105963
Description
Summary:A new, simple, specific, sensitive, rapid, accurate and precise RP-HPLC method was developed for the estimation of quetiapine in bulk and pharmaceutical formulations. Quetiapine was chromatographed on a reverse phase C18 Waters column (75x4.6 mm I.D., particle size 3.5 μm) in a mobile phase consisting of phosphate buffer (pH 3.0 adjusted with orthophosphoric acid) and acetonitrile in the ratio 40:60 v/v. The mobile phase was pumped at a flow rate of 0.8 mL/min with detection at 291 nm. The detector response was linear in the concentration of 20-120 μg/mL. The limit of detection and limit of quantitation was found to be 0.2 and 0.75 μg/mL, respectively. The intra and inter day variation was found to be less than 1%. The mean recovery of the drug from the solution was 99%. The proposed method is simple, fast, accurate, precise and reproducible hence, it can be applied for routine quality control analysis of quetiapine in bulk and pharmaceutical formulations.
ISSN:0973-4945
2090-9810