Factors affecting the minimum detected activity of the GAMMAVISION software report protocol

The article presents the results of a gamma- spectrometer study with a low- capacity scintillation (NaI(Tl)) detection unit, with software GammaVision v.8, adopted to solve certification problems, including for measuring activities close to zero. The list of factors influencing the presentation of t...

Full description

Bibliographic Details
Main Authors: T. Vinokurova, Z. Malimon, V. Salata, T. Prokopenko, G. Kochetova, L. Gusak
Format: Article
Language:English
Published: Stepan Gzhytskyi National University of Veterinary Medicine and Biotechnologies Lviv 2020-08-01
Series:Науковий вісник Львівського національного університету ветеринарної медицини та біотехнологій імені С.З. Ґжицького: Серія Ветеринарні науки
Subjects:
Online Access:https://nvlvet.com.ua/index.php/journal/article/view/3990
id doaj-1ff9e3ac63414da5b15821fa16807eec
record_format Article
spelling doaj-1ff9e3ac63414da5b15821fa16807eec2020-11-25T03:47:22ZengStepan Gzhytskyi National University of Veterinary Medicine and Biotechnologies LvivНауковий вісник Львівського національного університету ветеринарної медицини та біотехнологій імені С.З. Ґжицького: Серія Ветеринарні науки2518-75542518-13272020-08-01229813814310.32718/nvlvet98243990Factors affecting the minimum detected activity of the GAMMAVISION software report protocolT. Vinokurova0Z. Malimon1V. Salata2T. Prokopenko3G. Kochetova4L. Gusak5State Scientific Research Institute of Laboratory Diagnostics and Veterinary Sanitary Expertise, Kyiv, UkraineState Scientific Research Institute of Laboratory Diagnostics and Veterinary Sanitary Expertise, Kyiv, UkraineStepan Gzhytskyi National University of Veterinary Medicine and Biotechnologies Lviv, UkraineState Scientific Research Institute of Laboratory Diagnostics and Veterinary Sanitary Expertise, Kyiv, UkraineState Scientific Research Institute of Laboratory Diagnostics and Veterinary Sanitary Expertise, Kyiv, UkraineState Scientific Research Institute of Laboratory Diagnostics and Veterinary Sanitary Expertise, Kyiv, UkraineThe article presents the results of a gamma- spectrometer study with a low- capacity scintillation (NaI(Tl)) detection unit, with software GammaVision v.8, adopted to solve certification problems, including for measuring activities close to zero. The list of factors influencing the presentation of the minimum detected activity (MDA) in the report protocol is analyzed software GammaVision v.8. To perform the measurements, the company´s ORTEC gamma-spectrometer with a low-capacity detection  unit a type digiBASE-RH, in which a high-voltage power supply unit, amplifier, digital multi-channel analyzer with USB-connection to a computer with the installed driver are installed, was used. The peak search in the library was performed using a compressed version of the library, which contains two nuclides 137Cs and 40K. Methods of calculating MDA are based on the intensity of accounts. Nuclide activity is calculation for all peaks in the library, the energy of which is in the selected energy ranges for analysis. Measurements and calculations MDA under the given conditions of research for five methods depending on time of measurement and geometry of measurements are carried out. It is shown that the level of MDA can be reduced by choosing the optimal conditions, time of measurement and geometry of measurements. Recommendations are given for the use of capacity for counting sample, which will reduce the amount of substance required for analysis and optimize the time spent on measuring the counting sample. In a prospect it is expedient to undertake a study of dependence of MDA from the physical properties and radionuclide composition of the substance for the counting sample and the dependence from the uncertainty in the equation MDA.https://nvlvet.com.ua/index.php/journal/article/view/3990low capacity detection unit, mda, calculation methods, measurement conditions, measurement geometry.
collection DOAJ
language English
format Article
sources DOAJ
author T. Vinokurova
Z. Malimon
V. Salata
T. Prokopenko
G. Kochetova
L. Gusak
spellingShingle T. Vinokurova
Z. Malimon
V. Salata
T. Prokopenko
G. Kochetova
L. Gusak
Factors affecting the minimum detected activity of the GAMMAVISION software report protocol
Науковий вісник Львівського національного університету ветеринарної медицини та біотехнологій імені С.З. Ґжицького: Серія Ветеринарні науки
low capacity detection unit, mda, calculation methods, measurement conditions, measurement geometry.
author_facet T. Vinokurova
Z. Malimon
V. Salata
T. Prokopenko
G. Kochetova
L. Gusak
author_sort T. Vinokurova
title Factors affecting the minimum detected activity of the GAMMAVISION software report protocol
title_short Factors affecting the minimum detected activity of the GAMMAVISION software report protocol
title_full Factors affecting the minimum detected activity of the GAMMAVISION software report protocol
title_fullStr Factors affecting the minimum detected activity of the GAMMAVISION software report protocol
title_full_unstemmed Factors affecting the minimum detected activity of the GAMMAVISION software report protocol
title_sort factors affecting the minimum detected activity of the gammavision software report protocol
publisher Stepan Gzhytskyi National University of Veterinary Medicine and Biotechnologies Lviv
series Науковий вісник Львівського національного університету ветеринарної медицини та біотехнологій імені С.З. Ґжицького: Серія Ветеринарні науки
issn 2518-7554
2518-1327
publishDate 2020-08-01
description The article presents the results of a gamma- spectrometer study with a low- capacity scintillation (NaI(Tl)) detection unit, with software GammaVision v.8, adopted to solve certification problems, including for measuring activities close to zero. The list of factors influencing the presentation of the minimum detected activity (MDA) in the report protocol is analyzed software GammaVision v.8. To perform the measurements, the company´s ORTEC gamma-spectrometer with a low-capacity detection  unit a type digiBASE-RH, in which a high-voltage power supply unit, amplifier, digital multi-channel analyzer with USB-connection to a computer with the installed driver are installed, was used. The peak search in the library was performed using a compressed version of the library, which contains two nuclides 137Cs and 40K. Methods of calculating MDA are based on the intensity of accounts. Nuclide activity is calculation for all peaks in the library, the energy of which is in the selected energy ranges for analysis. Measurements and calculations MDA under the given conditions of research for five methods depending on time of measurement and geometry of measurements are carried out. It is shown that the level of MDA can be reduced by choosing the optimal conditions, time of measurement and geometry of measurements. Recommendations are given for the use of capacity for counting sample, which will reduce the amount of substance required for analysis and optimize the time spent on measuring the counting sample. In a prospect it is expedient to undertake a study of dependence of MDA from the physical properties and radionuclide composition of the substance for the counting sample and the dependence from the uncertainty in the equation MDA.
topic low capacity detection unit, mda, calculation methods, measurement conditions, measurement geometry.
url https://nvlvet.com.ua/index.php/journal/article/view/3990
work_keys_str_mv AT tvinokurova factorsaffectingtheminimumdetectedactivityofthegammavisionsoftwarereportprotocol
AT zmalimon factorsaffectingtheminimumdetectedactivityofthegammavisionsoftwarereportprotocol
AT vsalata factorsaffectingtheminimumdetectedactivityofthegammavisionsoftwarereportprotocol
AT tprokopenko factorsaffectingtheminimumdetectedactivityofthegammavisionsoftwarereportprotocol
AT gkochetova factorsaffectingtheminimumdetectedactivityofthegammavisionsoftwarereportprotocol
AT lgusak factorsaffectingtheminimumdetectedactivityofthegammavisionsoftwarereportprotocol
_version_ 1724502267600568320