Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500

AIM: To compare the differences and agreement in axial length(AL), anterior chamber depth(ACD)flat axial corneal curvature(K1)and steep axial corneal curvature(K2)measured by SW-9000 and IOL Master 500.METHODS: Totally 258 eyes of 129 patients with ametropia were measured with SW-9000 and IOL Master...

Full description

Bibliographic Details
Main Authors: Yan Li, Guo-Bao Li, Hui Shen
Format: Article
Language:English
Published: Press of International Journal of Ophthalmology (IJO PRESS) 2021-10-01
Series:Guoji Yanke Zazhi
Subjects:
Online Access:http://ies.ijo.cn/cn_publish/2021/10/202110025.pdf
id doaj-1efa985386614f559dca648ed9fbebd3
record_format Article
spelling doaj-1efa985386614f559dca648ed9fbebd32021-09-16T14:50:37ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51232021-10-0121101786179110.3980/j.issn.1672-5123.2021.10.25202110025Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500Yan Li0Guo-Bao Li1Hui Shen2Department of Ophthalmology, Sinopharm Dongfeng General Hospital, Hubei University of Medicine, Shiyan 442008, Hubei Province, ChinaDepartment of Ophthalmology, Sinopharm Dongfeng General Hospital, Hubei University of Medicine, Shiyan 442008, Hubei Province, ChinaDepartment of Ophthalmology, Sinopharm Dongfeng General Hospital, Hubei University of Medicine, Shiyan 442008, Hubei Province, ChinaAIM: To compare the differences and agreement in axial length(AL), anterior chamber depth(ACD)flat axial corneal curvature(K1)and steep axial corneal curvature(K2)measured by SW-9000 and IOL Master 500.METHODS: Totally 258 eyes of 129 patients with ametropia were measured with SW-9000 and IOL Master 500. Among them, 159 eyes were low to moderate myopia(-0.75D≤ spherical equivalent <-6D)and 99 eyes were high myopia(spherical equivalent ≥-6D). Comparison between two devices were performed for AL, ACD, K1, and K2. The difference between two instruments were evaluated with a paired samples t-test. The correlation between the measurements was evaluated using Pearson correlation coefficients. The agreement between the devices was evaluated with Bland-Altman plots. RESULTS: There was significant difference between SW-9000 and IOL Master 500 in measuring AL and ACD(all P<0.01), but there was no significant difference between K1 and K2 in patients with low to moderate myopia. There was no significant difference in AL, ACD and K1 between SW-9000 and IOL Master 500(all P>0.05), but there was significant difference in K2(P<0.05)in patients with high myopia. Pearson correlation analysis showed that AL, ACD, K1 and K2 of patients with low to moderate and high myopia were highly correlated(all P<0.01). Bland-Altman consistency analysis showed that the two methods had good consistency in measuring AL in patients with low to moderate and high myopia, but poor consistency in measuring ACD, K1 and K2.CONCLUSION: SW-9000 and IOL Master 500 have good consistency in measuring the AL of people with ametropia. The measurement of ACD and corneal curvature has poor consistency, and should be carefully selected in combination with clinical practice.http://ies.ijo.cn/cn_publish/2021/10/202110025.pdfsw-9000iol master 500axial lengthanterior chamber depthcorneal curvatureagreement
collection DOAJ
language English
format Article
sources DOAJ
author Yan Li
Guo-Bao Li
Hui Shen
spellingShingle Yan Li
Guo-Bao Li
Hui Shen
Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500
Guoji Yanke Zazhi
sw-9000
iol master 500
axial length
anterior chamber depth
corneal curvature
agreement
author_facet Yan Li
Guo-Bao Li
Hui Shen
author_sort Yan Li
title Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500
title_short Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500
title_full Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500
title_fullStr Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500
title_full_unstemmed Comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from SW-9000 and IOL Master 500
title_sort comparison of axial length, anterior chamber depth and corneal curvature parameters obtained from sw-9000 and iol master 500
publisher Press of International Journal of Ophthalmology (IJO PRESS)
series Guoji Yanke Zazhi
issn 1672-5123
publishDate 2021-10-01
description AIM: To compare the differences and agreement in axial length(AL), anterior chamber depth(ACD)flat axial corneal curvature(K1)and steep axial corneal curvature(K2)measured by SW-9000 and IOL Master 500.METHODS: Totally 258 eyes of 129 patients with ametropia were measured with SW-9000 and IOL Master 500. Among them, 159 eyes were low to moderate myopia(-0.75D≤ spherical equivalent <-6D)and 99 eyes were high myopia(spherical equivalent ≥-6D). Comparison between two devices were performed for AL, ACD, K1, and K2. The difference between two instruments were evaluated with a paired samples t-test. The correlation between the measurements was evaluated using Pearson correlation coefficients. The agreement between the devices was evaluated with Bland-Altman plots. RESULTS: There was significant difference between SW-9000 and IOL Master 500 in measuring AL and ACD(all P<0.01), but there was no significant difference between K1 and K2 in patients with low to moderate myopia. There was no significant difference in AL, ACD and K1 between SW-9000 and IOL Master 500(all P>0.05), but there was significant difference in K2(P<0.05)in patients with high myopia. Pearson correlation analysis showed that AL, ACD, K1 and K2 of patients with low to moderate and high myopia were highly correlated(all P<0.01). Bland-Altman consistency analysis showed that the two methods had good consistency in measuring AL in patients with low to moderate and high myopia, but poor consistency in measuring ACD, K1 and K2.CONCLUSION: SW-9000 and IOL Master 500 have good consistency in measuring the AL of people with ametropia. The measurement of ACD and corneal curvature has poor consistency, and should be carefully selected in combination with clinical practice.
topic sw-9000
iol master 500
axial length
anterior chamber depth
corneal curvature
agreement
url http://ies.ijo.cn/cn_publish/2021/10/202110025.pdf
work_keys_str_mv AT yanli comparisonofaxiallengthanteriorchamberdepthandcornealcurvatureparametersobtainedfromsw9000andiolmaster500
AT guobaoli comparisonofaxiallengthanteriorchamberdepthandcornealcurvatureparametersobtainedfromsw9000andiolmaster500
AT huishen comparisonofaxiallengthanteriorchamberdepthandcornealcurvatureparametersobtainedfromsw9000andiolmaster500
_version_ 1717378127692300288