Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
Rapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central labora...
Main Authors: | , , , , , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Public Library of Science (PLoS)
2017-01-01
|
Series: | PLoS ONE |
Online Access: | http://europepmc.org/articles/PMC5224825?pdf=render |
id |
doaj-1a142a8c18034afb9c7ac86d9214c733 |
---|---|
record_format |
Article |
spelling |
doaj-1a142a8c18034afb9c7ac86d9214c7332020-11-24T20:50:01ZengPublic Library of Science (PLoS)PLoS ONE1932-62032017-01-01121e016959310.1371/journal.pone.0169593Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.Jérôme Allardet-ServentMelissa LebsirChristian DubrocaMartine FabrigouleSylvie JordanaThomas SignouretMatthias CastanierGuillemette ThomasRettinavelou SoundaravelouAnne LepidiLaurence DelapierreGuillaume PenarandaPhilippe HalfonJean-Marie SeghboyanRapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central laboratory measurements of electrolytes, bicarbonate, hemoglobin, hematocrit, and glucose.314 paired samples were collected prospectively from 51 critically ill patients. All samples were drawn simultaneously in the morning from an arterial line. BD Vacutainer tubes were analyzed in the central laboratory using Beckman Coulter analyzers (AU 5800 and DxH 800). BD Preset 3 ml heparinized-syringes were analyzed immediately in the ICU using the POC Siemens RAPIDPoint 500 blood gas system. We used CLIA proficiency testing criteria to define acceptable analytical performance and interchangeability.Biases, limits of agreement (±1.96 SD) and coefficients of correlation were respectively: 1.3 (-2.2 to 4.8 mmol/L, r = 0.936) for sodium; 0.2 (-0.2 to 0.6 mmol/L, r = 0.944) for potassium; -0.9 (-3.7 to 2 mmol/L, r = 0.967) for chloride; 0.8 (-1.9 to 3.4 mmol/L, r = 0.968) for bicarbonate; -11 (-30 to 9 mg/dL, r = 0.972) for glucose; -0.8 (-1.4 to -0.2 g/dL, r = 0.985) for hemoglobin; and -1.1 (-2.9 to 0.7%, r = 0.981) for hematocrit. All differences were below CLIA cut-off values, except for hemoglobin.Compared to central Laboratory analyzers, the POC Siemens RAPIDPoint 500 blood gas system satisfied the CLIA criteria of interchangeability for all tested parameters, except for hemoglobin. These results are warranted for our own procedures and devices. Bearing these restrictions, we recommend clinicians to initiate an appropriate therapy based on POC testing without awaiting a control measurement.http://europepmc.org/articles/PMC5224825?pdf=render |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Jérôme Allardet-Servent Melissa Lebsir Christian Dubroca Martine Fabrigoule Sylvie Jordana Thomas Signouret Matthias Castanier Guillemette Thomas Rettinavelou Soundaravelou Anne Lepidi Laurence Delapierre Guillaume Penaranda Philippe Halfon Jean-Marie Seghboyan |
spellingShingle |
Jérôme Allardet-Servent Melissa Lebsir Christian Dubroca Martine Fabrigoule Sylvie Jordana Thomas Signouret Matthias Castanier Guillemette Thomas Rettinavelou Soundaravelou Anne Lepidi Laurence Delapierre Guillaume Penaranda Philippe Halfon Jean-Marie Seghboyan Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients. PLoS ONE |
author_facet |
Jérôme Allardet-Servent Melissa Lebsir Christian Dubroca Martine Fabrigoule Sylvie Jordana Thomas Signouret Matthias Castanier Guillemette Thomas Rettinavelou Soundaravelou Anne Lepidi Laurence Delapierre Guillaume Penaranda Philippe Halfon Jean-Marie Seghboyan |
author_sort |
Jérôme Allardet-Servent |
title |
Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients. |
title_short |
Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients. |
title_full |
Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients. |
title_fullStr |
Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients. |
title_full_unstemmed |
Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients. |
title_sort |
point-of-care versus central laboratory measurements of hemoglobin, hematocrit, glucose, bicarbonate and electrolytes: a prospective observational study in critically ill patients. |
publisher |
Public Library of Science (PLoS) |
series |
PLoS ONE |
issn |
1932-6203 |
publishDate |
2017-01-01 |
description |
Rapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central laboratory measurements of electrolytes, bicarbonate, hemoglobin, hematocrit, and glucose.314 paired samples were collected prospectively from 51 critically ill patients. All samples were drawn simultaneously in the morning from an arterial line. BD Vacutainer tubes were analyzed in the central laboratory using Beckman Coulter analyzers (AU 5800 and DxH 800). BD Preset 3 ml heparinized-syringes were analyzed immediately in the ICU using the POC Siemens RAPIDPoint 500 blood gas system. We used CLIA proficiency testing criteria to define acceptable analytical performance and interchangeability.Biases, limits of agreement (±1.96 SD) and coefficients of correlation were respectively: 1.3 (-2.2 to 4.8 mmol/L, r = 0.936) for sodium; 0.2 (-0.2 to 0.6 mmol/L, r = 0.944) for potassium; -0.9 (-3.7 to 2 mmol/L, r = 0.967) for chloride; 0.8 (-1.9 to 3.4 mmol/L, r = 0.968) for bicarbonate; -11 (-30 to 9 mg/dL, r = 0.972) for glucose; -0.8 (-1.4 to -0.2 g/dL, r = 0.985) for hemoglobin; and -1.1 (-2.9 to 0.7%, r = 0.981) for hematocrit. All differences were below CLIA cut-off values, except for hemoglobin.Compared to central Laboratory analyzers, the POC Siemens RAPIDPoint 500 blood gas system satisfied the CLIA criteria of interchangeability for all tested parameters, except for hemoglobin. These results are warranted for our own procedures and devices. Bearing these restrictions, we recommend clinicians to initiate an appropriate therapy based on POC testing without awaiting a control measurement. |
url |
http://europepmc.org/articles/PMC5224825?pdf=render |
work_keys_str_mv |
AT jeromeallardetservent pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT melissalebsir pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT christiandubroca pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT martinefabrigoule pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT sylviejordana pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT thomassignouret pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT matthiascastanier pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT guillemettethomas pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT rettinavelousoundaravelou pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT annelepidi pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT laurencedelapierre pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT guillaumepenaranda pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT philippehalfon pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients AT jeanmarieseghboyan pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients |