Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.

Rapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central labora...

Full description

Bibliographic Details
Main Authors: Jérôme Allardet-Servent, Melissa Lebsir, Christian Dubroca, Martine Fabrigoule, Sylvie Jordana, Thomas Signouret, Matthias Castanier, Guillemette Thomas, Rettinavelou Soundaravelou, Anne Lepidi, Laurence Delapierre, Guillaume Penaranda, Philippe Halfon, Jean-Marie Seghboyan
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2017-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC5224825?pdf=render
id doaj-1a142a8c18034afb9c7ac86d9214c733
record_format Article
spelling doaj-1a142a8c18034afb9c7ac86d9214c7332020-11-24T20:50:01ZengPublic Library of Science (PLoS)PLoS ONE1932-62032017-01-01121e016959310.1371/journal.pone.0169593Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.Jérôme Allardet-ServentMelissa LebsirChristian DubrocaMartine FabrigouleSylvie JordanaThomas SignouretMatthias CastanierGuillemette ThomasRettinavelou SoundaravelouAnne LepidiLaurence DelapierreGuillaume PenarandaPhilippe HalfonJean-Marie SeghboyanRapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central laboratory measurements of electrolytes, bicarbonate, hemoglobin, hematocrit, and glucose.314 paired samples were collected prospectively from 51 critically ill patients. All samples were drawn simultaneously in the morning from an arterial line. BD Vacutainer tubes were analyzed in the central laboratory using Beckman Coulter analyzers (AU 5800 and DxH 800). BD Preset 3 ml heparinized-syringes were analyzed immediately in the ICU using the POC Siemens RAPIDPoint 500 blood gas system. We used CLIA proficiency testing criteria to define acceptable analytical performance and interchangeability.Biases, limits of agreement (±1.96 SD) and coefficients of correlation were respectively: 1.3 (-2.2 to 4.8 mmol/L, r = 0.936) for sodium; 0.2 (-0.2 to 0.6 mmol/L, r = 0.944) for potassium; -0.9 (-3.7 to 2 mmol/L, r = 0.967) for chloride; 0.8 (-1.9 to 3.4 mmol/L, r = 0.968) for bicarbonate; -11 (-30 to 9 mg/dL, r = 0.972) for glucose; -0.8 (-1.4 to -0.2 g/dL, r = 0.985) for hemoglobin; and -1.1 (-2.9 to 0.7%, r = 0.981) for hematocrit. All differences were below CLIA cut-off values, except for hemoglobin.Compared to central Laboratory analyzers, the POC Siemens RAPIDPoint 500 blood gas system satisfied the CLIA criteria of interchangeability for all tested parameters, except for hemoglobin. These results are warranted for our own procedures and devices. Bearing these restrictions, we recommend clinicians to initiate an appropriate therapy based on POC testing without awaiting a control measurement.http://europepmc.org/articles/PMC5224825?pdf=render
collection DOAJ
language English
format Article
sources DOAJ
author Jérôme Allardet-Servent
Melissa Lebsir
Christian Dubroca
Martine Fabrigoule
Sylvie Jordana
Thomas Signouret
Matthias Castanier
Guillemette Thomas
Rettinavelou Soundaravelou
Anne Lepidi
Laurence Delapierre
Guillaume Penaranda
Philippe Halfon
Jean-Marie Seghboyan
spellingShingle Jérôme Allardet-Servent
Melissa Lebsir
Christian Dubroca
Martine Fabrigoule
Sylvie Jordana
Thomas Signouret
Matthias Castanier
Guillemette Thomas
Rettinavelou Soundaravelou
Anne Lepidi
Laurence Delapierre
Guillaume Penaranda
Philippe Halfon
Jean-Marie Seghboyan
Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
PLoS ONE
author_facet Jérôme Allardet-Servent
Melissa Lebsir
Christian Dubroca
Martine Fabrigoule
Sylvie Jordana
Thomas Signouret
Matthias Castanier
Guillemette Thomas
Rettinavelou Soundaravelou
Anne Lepidi
Laurence Delapierre
Guillaume Penaranda
Philippe Halfon
Jean-Marie Seghboyan
author_sort Jérôme Allardet-Servent
title Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
title_short Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
title_full Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
title_fullStr Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
title_full_unstemmed Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
title_sort point-of-care versus central laboratory measurements of hemoglobin, hematocrit, glucose, bicarbonate and electrolytes: a prospective observational study in critically ill patients.
publisher Public Library of Science (PLoS)
series PLoS ONE
issn 1932-6203
publishDate 2017-01-01
description Rapid detection of abnormal biological values using point-of-care (POC) testing allows clinicians to promptly initiate therapy; however, there are concerns regarding the reliability of POC measurements. We investigated the agreement between the latest generation blood gas analyzer and central laboratory measurements of electrolytes, bicarbonate, hemoglobin, hematocrit, and glucose.314 paired samples were collected prospectively from 51 critically ill patients. All samples were drawn simultaneously in the morning from an arterial line. BD Vacutainer tubes were analyzed in the central laboratory using Beckman Coulter analyzers (AU 5800 and DxH 800). BD Preset 3 ml heparinized-syringes were analyzed immediately in the ICU using the POC Siemens RAPIDPoint 500 blood gas system. We used CLIA proficiency testing criteria to define acceptable analytical performance and interchangeability.Biases, limits of agreement (±1.96 SD) and coefficients of correlation were respectively: 1.3 (-2.2 to 4.8 mmol/L, r = 0.936) for sodium; 0.2 (-0.2 to 0.6 mmol/L, r = 0.944) for potassium; -0.9 (-3.7 to 2 mmol/L, r = 0.967) for chloride; 0.8 (-1.9 to 3.4 mmol/L, r = 0.968) for bicarbonate; -11 (-30 to 9 mg/dL, r = 0.972) for glucose; -0.8 (-1.4 to -0.2 g/dL, r = 0.985) for hemoglobin; and -1.1 (-2.9 to 0.7%, r = 0.981) for hematocrit. All differences were below CLIA cut-off values, except for hemoglobin.Compared to central Laboratory analyzers, the POC Siemens RAPIDPoint 500 blood gas system satisfied the CLIA criteria of interchangeability for all tested parameters, except for hemoglobin. These results are warranted for our own procedures and devices. Bearing these restrictions, we recommend clinicians to initiate an appropriate therapy based on POC testing without awaiting a control measurement.
url http://europepmc.org/articles/PMC5224825?pdf=render
work_keys_str_mv AT jeromeallardetservent pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT melissalebsir pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT christiandubroca pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT martinefabrigoule pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT sylviejordana pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT thomassignouret pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT matthiascastanier pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT guillemettethomas pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT rettinavelousoundaravelou pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT annelepidi pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT laurencedelapierre pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT guillaumepenaranda pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT philippehalfon pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
AT jeanmarieseghboyan pointofcareversuscentrallaboratorymeasurementsofhemoglobinhematocritglucosebicarbonateandelectrolytesaprospectiveobservationalstudyincriticallyillpatients
_version_ 1716805056799440896