Metrological Array of Cyber-Physical Systems. Part 11. Remote Error Correction of Measuring Channel

The multi-channel measuring instruments with both the classical structure and the isolated one is identified their errors major factors basing on general it metrological properties analysis. Limiting possibilities of the remote automatic method for additive and multiplicative errors correction of me...

Full description

Bibliographic Details
Main Authors: Yuriy YATSUK, Mykola MYKYJCHUK, Volodymyr ZDEB, Roman YANOVYCH
Format: Article
Language:English
Published: IFSA Publishing, S.L. 2015-09-01
Series:Sensors & Transducers
Subjects:
Online Access:http://www.sensorsportal.com/HTML/DIGEST/september_2015/Vol_192/P_2719.pdf
Description
Summary:The multi-channel measuring instruments with both the classical structure and the isolated one is identified their errors major factors basing on general it metrological properties analysis. Limiting possibilities of the remote automatic method for additive and multiplicative errors correction of measuring instruments with help of code-control measures are studied. For on-site calibration of multi- channel measuring instruments, the portable voltage calibrators structures are suggested and their metrological properties while automatic errors adjusting are analysed. It was experimentally envisaged that unadjusted error value does not exceed ± 1 mV that satisfies most industrial applications. This has confirmed the main approval concerning the possibilities of remote errors self-adjustment as well multi- channel measuring instruments as calibration tools for proper verification.