Thermal modelling of the calorimeter used for energy measurement of LMJ laser pulse
In this work, a 3D thermal model of the LMJ calorimeter is developed using Comsol multiphysics. The unknown thermal properties of the system are identified by comparing the model results with the measured temperature. Several model's applications, such as comparing deposition modes (electrical...
Main Authors: | Crespy C., Villate D. |
---|---|
Format: | Article |
Language: | English |
Published: |
EDP Sciences
2013-11-01
|
Series: | EPJ Web of Conferences |
Online Access: | http://dx.doi.org/10.1051/epjconf/20135908006 |