Summary: | An alloy surface with possible applications in the medical field, Ti4A14Zr, was improved through the deposition of a thin hydroxyapatite (HA) layer. In this paper, we analyzed the growth of a HA layer through electrophoresis and the corrosion resistance of the metallic sample covered with the ceramic layer. The substrate surface was processed via chemical procedures before the HA deposition. The state of the metallic surface and that of the layer of HA were investigated using scanning electron microscopy (SEM) and energy dispersive spectroscopy (EDS) analysis of the chemical composition. The results indicate a high increase in the corrosion resistance associated with the ceramic layer compared to the metallic basic layer. Moreover, the analysis revealed the formation of a homogeneous TiO<sub>2</sub> layer on the surface of the metallic substrate. The titanium oxide layer identified by SEM–EDS and confirmed by EIS was very homogeneous and resistant, with a compact microstructural appearance and submicron dimension. The layer composed of TiO<sub>2</sub> and HA provided good corrosion protection.
|