Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies

Excessive imaging dose from repeated scans and poor image quality mainly due to scatter contamination are the two bottlenecks of cone-beam CT (CBCT) imaging. Compressed sensing (CS) reconstruction algorithms show promises in recovering faithful signals from low-dose projection data but do not serve...

Full description

Bibliographic Details
Main Authors: Xue Dong, Michael Petrongolo, Tianye Niu, Lei Zhu
Format: Article
Language:English
Published: Hindawi Limited 2013-01-01
Series:Computational and Mathematical Methods in Medicine
Online Access:http://dx.doi.org/10.1155/2013/637614
id doaj-0ffe6a3fde8040c4bc6ee707ad07ac91
record_format Article
spelling doaj-0ffe6a3fde8040c4bc6ee707ad07ac912020-11-24T21:05:56ZengHindawi LimitedComputational and Mathematical Methods in Medicine1748-670X1748-67182013-01-01201310.1155/2013/637614637614Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom StudiesXue Dong0Michael Petrongolo1Tianye Niu2Lei Zhu3Nuclear & Radiological Engineering and Medical Physics Programs, The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USANuclear & Radiological Engineering and Medical Physics Programs, The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USANuclear & Radiological Engineering and Medical Physics Programs, The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USANuclear & Radiological Engineering and Medical Physics Programs, The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USAExcessive imaging dose from repeated scans and poor image quality mainly due to scatter contamination are the two bottlenecks of cone-beam CT (CBCT) imaging. Compressed sensing (CS) reconstruction algorithms show promises in recovering faithful signals from low-dose projection data but do not serve well the needs of accurate CBCT imaging if effective scatter correction is not in place. Scatter can be accurately measured and removed using measurement-based methods. However, these approaches are considered unpractical in the conventional FDK reconstruction, due to the inevitable primary loss for scatter measurement. We combine measurement-based scatter correction and CS-based iterative reconstruction to generate scatter-free images from low-dose projections. We distribute blocked areas on the detector where primary signals are considered redundant in a full scan. Scatter distribution is estimated by interpolating/extrapolating measured scatter samples inside blocked areas. CS-based iterative reconstruction is finally carried out on the undersampled data to obtain scatter-free and low-dose CBCT images. With only 25% of conventional full-scan dose, our method reduces the average CT number error from 250 HU to 24 HU and increases the contrast by a factor of 2.1 on Catphan 600 phantom. On an anthropomorphic head phantom, the average CT number error is reduced from 224 HU to 10 HU in the central uniform area.http://dx.doi.org/10.1155/2013/637614
collection DOAJ
language English
format Article
sources DOAJ
author Xue Dong
Michael Petrongolo
Tianye Niu
Lei Zhu
spellingShingle Xue Dong
Michael Petrongolo
Tianye Niu
Lei Zhu
Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies
Computational and Mathematical Methods in Medicine
author_facet Xue Dong
Michael Petrongolo
Tianye Niu
Lei Zhu
author_sort Xue Dong
title Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies
title_short Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies
title_full Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies
title_fullStr Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies
title_full_unstemmed Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies
title_sort low-dose and scatter-free cone-beam ct imaging using a stationary beam blocker in a single scan: phantom studies
publisher Hindawi Limited
series Computational and Mathematical Methods in Medicine
issn 1748-670X
1748-6718
publishDate 2013-01-01
description Excessive imaging dose from repeated scans and poor image quality mainly due to scatter contamination are the two bottlenecks of cone-beam CT (CBCT) imaging. Compressed sensing (CS) reconstruction algorithms show promises in recovering faithful signals from low-dose projection data but do not serve well the needs of accurate CBCT imaging if effective scatter correction is not in place. Scatter can be accurately measured and removed using measurement-based methods. However, these approaches are considered unpractical in the conventional FDK reconstruction, due to the inevitable primary loss for scatter measurement. We combine measurement-based scatter correction and CS-based iterative reconstruction to generate scatter-free images from low-dose projections. We distribute blocked areas on the detector where primary signals are considered redundant in a full scan. Scatter distribution is estimated by interpolating/extrapolating measured scatter samples inside blocked areas. CS-based iterative reconstruction is finally carried out on the undersampled data to obtain scatter-free and low-dose CBCT images. With only 25% of conventional full-scan dose, our method reduces the average CT number error from 250 HU to 24 HU and increases the contrast by a factor of 2.1 on Catphan 600 phantom. On an anthropomorphic head phantom, the average CT number error is reduced from 224 HU to 10 HU in the central uniform area.
url http://dx.doi.org/10.1155/2013/637614
work_keys_str_mv AT xuedong lowdoseandscatterfreeconebeamctimagingusingastationarybeamblockerinasinglescanphantomstudies
AT michaelpetrongolo lowdoseandscatterfreeconebeamctimagingusingastationarybeamblockerinasinglescanphantomstudies
AT tianyeniu lowdoseandscatterfreeconebeamctimagingusingastationarybeamblockerinasinglescanphantomstudies
AT leizhu lowdoseandscatterfreeconebeamctimagingusingastationarybeamblockerinasinglescanphantomstudies
_version_ 1716767393195229184