The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
Background: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2014-01-01
|
Series: | Dental Research Journal |
Subjects: | |
Online Access: | http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=1;spage=74;epage=80;aulast=Fayaz |
id |
doaj-0f927a480f114c7588eb7efcae678014 |
---|---|
record_format |
Article |
spelling |
doaj-0f927a480f114c7588eb7efcae6780142020-11-25T00:19:04ZengWolters Kluwer Medknow PublicationsDental Research Journal1735-33272008-02552014-01-01111748010.4103/1735-3327.127193The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implantsAli FayazMinoo MahshidAboulfazl SabourySeyed Jalil SadrGhassem AnsariBackground: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values. Materials and Methods: Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05). Results: The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05). Conclusion: The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles.http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=1;spage=74;epage=80;aulast=FayazCalibrationdental implantsequipment failuretorque |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Ali Fayaz Minoo Mahshid Aboulfazl Saboury Seyed Jalil Sadr Ghassem Ansari |
spellingShingle |
Ali Fayaz Minoo Mahshid Aboulfazl Saboury Seyed Jalil Sadr Ghassem Ansari The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants Dental Research Journal Calibration dental implants equipment failure torque |
author_facet |
Ali Fayaz Minoo Mahshid Aboulfazl Saboury Seyed Jalil Sadr Ghassem Ansari |
author_sort |
Ali Fayaz |
title |
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants |
title_short |
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants |
title_full |
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants |
title_fullStr |
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants |
title_full_unstemmed |
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants |
title_sort |
effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants |
publisher |
Wolters Kluwer Medknow Publications |
series |
Dental Research Journal |
issn |
1735-3327 2008-0255 |
publishDate |
2014-01-01 |
description |
Background: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values.
Materials and Methods: Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05).
Results: The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05).
Conclusion: The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles. |
topic |
Calibration dental implants equipment failure torque |
url |
http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=1;spage=74;epage=80;aulast=Fayaz |
work_keys_str_mv |
AT alifayaz theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT minoomahshid theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT aboulfazlsaboury theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT seyedjalilsadr theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT ghassemansari theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT alifayaz effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT minoomahshid effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT aboulfazlsaboury effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT seyedjalilsadr effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants AT ghassemansari effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants |
_version_ |
1725373382034194432 |