The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants

Background: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of...

Full description

Bibliographic Details
Main Authors: Ali Fayaz, Minoo Mahshid, Aboulfazl Saboury, Seyed Jalil Sadr, Ghassem Ansari
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2014-01-01
Series:Dental Research Journal
Subjects:
Online Access:http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=1;spage=74;epage=80;aulast=Fayaz
id doaj-0f927a480f114c7588eb7efcae678014
record_format Article
spelling doaj-0f927a480f114c7588eb7efcae6780142020-11-25T00:19:04ZengWolters Kluwer Medknow PublicationsDental Research Journal1735-33272008-02552014-01-01111748010.4103/1735-3327.127193The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implantsAli FayazMinoo MahshidAboulfazl SabourySeyed Jalil SadrGhassem AnsariBackground: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values. Materials and Methods: Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05). Results: The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05). Conclusion: The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles.http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=1;spage=74;epage=80;aulast=FayazCalibrationdental implantsequipment failuretorque
collection DOAJ
language English
format Article
sources DOAJ
author Ali Fayaz
Minoo Mahshid
Aboulfazl Saboury
Seyed Jalil Sadr
Ghassem Ansari
spellingShingle Ali Fayaz
Minoo Mahshid
Aboulfazl Saboury
Seyed Jalil Sadr
Ghassem Ansari
The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
Dental Research Journal
Calibration
dental implants
equipment failure
torque
author_facet Ali Fayaz
Minoo Mahshid
Aboulfazl Saboury
Seyed Jalil Sadr
Ghassem Ansari
author_sort Ali Fayaz
title The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_short The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_full The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_fullStr The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_full_unstemmed The effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
title_sort effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices for dental implants
publisher Wolters Kluwer Medknow Publications
series Dental Research Journal
issn 1735-3327
2008-0255
publishDate 2014-01-01
description Background: Mechanical torque limiting devices (MTLDs) are necessary tools to control a peak torque and achieving target values of screw component of dental implants. Due to probable effect of autoclaving and number of use on the accuracy of these devices, this study aimed to evaluate the effect of sterilization and number of use on the accuracy of friction-style mechanical torque limiting devices (F-S MTLDs) in achieving their target torque values. Materials and Methods: Peak torque measurements of 15 new F-S MTLDs from three different manufacturers (Astra Tech, BioHorizons, Dr. Idhe) were measured ten times before and after 100 steam sterilization using a digital torque gauge. To simulate the clinical situation of aging (number of use) target torque application process was repeated 10 times after each sterilization cycle and the peak torque values were registered. Comparison of the mean differences with target torque in each cycle was performed using one sample t test. Considering the type of MTLDs as inter subject comparison, One-way repeated measure ANOVA was used to evaluate the absolute values of differences between devices of each manufacturer in each group (α = 0.05). Results: The results of this study in Dr. Idhe group showed that, mean of difference values significantly differed from the target torque (P = 0.002) until 75 cycles. In Astra Tech group, also mean of difference values with under estimation trend, showed a significant difference with the target torque (P < 0.001). Mean of difference values significantly differed from the target torque with under estimation trend during all the 100 cycles in BioHorizons group (P < 0.05). Conclusion: The torque output of each individual device stayed in 10% difference from target torque values before 100 sterilization cycles, but more than 10% difference from the target torque was seen in varying degrees during these consequent cycles.
topic Calibration
dental implants
equipment failure
torque
url http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=1;spage=74;epage=80;aulast=Fayaz
work_keys_str_mv AT alifayaz theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT minoomahshid theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT aboulfazlsaboury theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT seyedjalilsadr theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT ghassemansari theeffectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT alifayaz effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT minoomahshid effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT aboulfazlsaboury effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT seyedjalilsadr effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
AT ghassemansari effectofsterilizationandnumberofuseontheaccuracyoffrictionstylemechanicaltorquelimitingdevicesfordentalimplants
_version_ 1725373382034194432