2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene

Background and purpose — The long-term survival of arthroplasty components may be limited by polyethylene wear-related problems such as periprosthetic osteolysis and aseptic loosening. Highly cross-linked polyethylene (HXLPE) blended with vitamin E was introduced to improve oxidative stability and t...

Full description

Bibliographic Details
Main Authors: Joost H J van Erp, Julie R A Massier, Jelle J Halma, Thom E Snijders, Arthur de Gast
Format: Article
Language:English
Published: Taylor & Francis Group 2020-05-01
Series:Acta Orthopaedica
Online Access:http://dx.doi.org/10.1080/17453674.2020.1730073
id doaj-0eeb4bfa10f64e1b80dd6be424b54ac6
record_format Article
spelling doaj-0eeb4bfa10f64e1b80dd6be424b54ac62021-04-02T13:40:08ZengTaylor & Francis GroupActa Orthopaedica1745-36741745-36822020-05-0191325425910.1080/17453674.2020.173007317300732-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethyleneJoost H J van Erp0Julie R A Massier1Jelle J Halma2Thom E Snijders3Arthur de Gast4Clinical Orthopedic Research Center—mNClinical Orthopedic Research Center—mNClinical Orthopedic Research Center—mNClinical Orthopedic Research Center—mNClinical Orthopedic Research Center—mNBackground and purpose — The long-term survival of arthroplasty components may be limited by polyethylene wear-related problems such as periprosthetic osteolysis and aseptic loosening. Highly cross-linked polyethylene (HXLPE) blended with vitamin E was introduced to improve oxidative stability and to avoid long-term embrittlement. This study clinically compares the tribological behavior and clinical outcome of vitamin E blended HXLPE with ultra-high molecular weight polyethylene (UHMWPE) in an isoelastic monoblock cup for uncemented total hip arthroplasty. Patients and methods — In this randomized controlled trial (RCT), 199 patients were included: 102 patients received the vitamin E blended HXLPE cup, 97 patients the UHMWPE cup. Clinical and radiographic parameters were obtained preoperatively, directly postoperative and at 3, 12, and 24 months. Wear rates were compared using the mean linear femoral head penetration (FHP) rate. Results — 188 patients (94%) completed the 2-year follow-up. Mean patient satisfaction was higher in the vitamin E blended HXLPE group (8.9 [1]) than in in the control group (8.5 [2], p = 0.03). The Harris Hip Score (HHS) was higher in the vitamin E blended HXLPE group (95 [8]) than in the control group (92 [11], p = 0.3). The FHP rate was lower in the vitamin E blended HXLPE group: 0.046 mm/year compared with 0.056 mm/year in the control group (p = 0.05). No adverse reactions associated with the clinical application of vitamin E blended HXLPE were observed during follow-up, with an excellent 2-year survival to revision rate of 98% for both cups. Interpretation — This study shows the superior performance of the HXLPE blended with vitamin E acetabular cup with lower linear femoral head penetration rates and better clinical results compared with the UHMWPE acetabular cup after 2 years.http://dx.doi.org/10.1080/17453674.2020.1730073
collection DOAJ
language English
format Article
sources DOAJ
author Joost H J van Erp
Julie R A Massier
Jelle J Halma
Thom E Snijders
Arthur de Gast
spellingShingle Joost H J van Erp
Julie R A Massier
Jelle J Halma
Thom E Snijders
Arthur de Gast
2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
Acta Orthopaedica
author_facet Joost H J van Erp
Julie R A Massier
Jelle J Halma
Thom E Snijders
Arthur de Gast
author_sort Joost H J van Erp
title 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
title_short 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
title_full 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
title_fullStr 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
title_full_unstemmed 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
title_sort 2-year results of an rct of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin e blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene
publisher Taylor & Francis Group
series Acta Orthopaedica
issn 1745-3674
1745-3682
publishDate 2020-05-01
description Background and purpose — The long-term survival of arthroplasty components may be limited by polyethylene wear-related problems such as periprosthetic osteolysis and aseptic loosening. Highly cross-linked polyethylene (HXLPE) blended with vitamin E was introduced to improve oxidative stability and to avoid long-term embrittlement. This study clinically compares the tribological behavior and clinical outcome of vitamin E blended HXLPE with ultra-high molecular weight polyethylene (UHMWPE) in an isoelastic monoblock cup for uncemented total hip arthroplasty. Patients and methods — In this randomized controlled trial (RCT), 199 patients were included: 102 patients received the vitamin E blended HXLPE cup, 97 patients the UHMWPE cup. Clinical and radiographic parameters were obtained preoperatively, directly postoperative and at 3, 12, and 24 months. Wear rates were compared using the mean linear femoral head penetration (FHP) rate. Results — 188 patients (94%) completed the 2-year follow-up. Mean patient satisfaction was higher in the vitamin E blended HXLPE group (8.9 [1]) than in in the control group (8.5 [2], p = 0.03). The Harris Hip Score (HHS) was higher in the vitamin E blended HXLPE group (95 [8]) than in the control group (92 [11], p = 0.3). The FHP rate was lower in the vitamin E blended HXLPE group: 0.046 mm/year compared with 0.056 mm/year in the control group (p = 0.05). No adverse reactions associated with the clinical application of vitamin E blended HXLPE were observed during follow-up, with an excellent 2-year survival to revision rate of 98% for both cups. Interpretation — This study shows the superior performance of the HXLPE blended with vitamin E acetabular cup with lower linear femoral head penetration rates and better clinical results compared with the UHMWPE acetabular cup after 2 years.
url http://dx.doi.org/10.1080/17453674.2020.1730073
work_keys_str_mv AT joosthjvanerp 2yearresultsofanrctof2uncementedisoelasticmonoblockacetabularcomponentslowerwearratewithvitamineblendedhighlycrosslinkedpolyethylenecomparedtoultrahighmolecularweightpolyethylene
AT julieramassier 2yearresultsofanrctof2uncementedisoelasticmonoblockacetabularcomponentslowerwearratewithvitamineblendedhighlycrosslinkedpolyethylenecomparedtoultrahighmolecularweightpolyethylene
AT jellejhalma 2yearresultsofanrctof2uncementedisoelasticmonoblockacetabularcomponentslowerwearratewithvitamineblendedhighlycrosslinkedpolyethylenecomparedtoultrahighmolecularweightpolyethylene
AT thomesnijders 2yearresultsofanrctof2uncementedisoelasticmonoblockacetabularcomponentslowerwearratewithvitamineblendedhighlycrosslinkedpolyethylenecomparedtoultrahighmolecularweightpolyethylene
AT arthurdegast 2yearresultsofanrctof2uncementedisoelasticmonoblockacetabularcomponentslowerwearratewithvitamineblendedhighlycrosslinkedpolyethylenecomparedtoultrahighmolecularweightpolyethylene
_version_ 1721564225885700096