Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance

<p>Abstract</p> <p>Background</p> <p>One potential source of error in phase contrast (PC) congenital CMR flow measurements is caused by phase offsets due to local non-compensated eddy currents. Phantom correction of these phase offset errors has been shown to result in...

Full description

Bibliographic Details
Main Authors: Lai Wyman W, Printz Beth F, Holland Brian J
Format: Article
Language:English
Published: BMC 2010-03-01
Series:Journal of Cardiovascular Magnetic Resonance
Online Access:http://www.jcmr-online.com/content/12/1/11
id doaj-088bfde3d5044957b6630f5420a0c02f
record_format Article
spelling doaj-088bfde3d5044957b6630f5420a0c02f2020-11-24T22:03:10ZengBMCJournal of Cardiovascular Magnetic Resonance1097-66471532-429X2010-03-011211110.1186/1532-429X-12-11Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonanceLai Wyman WPrintz Beth FHolland Brian J<p>Abstract</p> <p>Background</p> <p>One potential source of error in phase contrast (PC) congenital CMR flow measurements is caused by phase offsets due to local non-compensated eddy currents. Phantom correction of these phase offset errors has been shown to result in more accurate measurements of blood flow in adults with structurally normal hearts. We report the effect of phantom correction on PC flow measurements at a clinical congenital CMR program.</p> <p>Results</p> <p>Flow was measured in the ascending aorta, main pulmonary artery, and right and left pulmonary arteries as clinically indicated, and additional values such as Qp/Qs were derived from these measurements. Phantom correction in our study population of 149 patients resulted in clinically significant changes in 13% to 48% of these phase-contrast measurements in patients with known or suspected heart disease. Overall, 640 measurements or calculated values were analyzed, and clinically significant changes were found in 31%. Larger vessels were associated with greater phase offset errors, with 22% of the changes in PC flow measurements attributed to the size of the vessel measured. In patients with structurally normal hearts, the pulmonary-to-systemic flow ratio after phantom correction was closer to 1.0 than before phantom correction. There was no significant difference in the effect of phantom correction for patients with tetralogy of Fallot as compared to the group as a whole.</p> <p>Conclusions</p> <p>Phantom correction often resulted in clinically significant changes in PC blood flow measurements in patients with known or suspected congenital heart disease. In laboratories performing clinical CMR with suspected phase offset errors of significance, the routine use of phantom correction for PC flow measurements should be considered.</p> http://www.jcmr-online.com/content/12/1/11
collection DOAJ
language English
format Article
sources DOAJ
author Lai Wyman W
Printz Beth F
Holland Brian J
spellingShingle Lai Wyman W
Printz Beth F
Holland Brian J
Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
Journal of Cardiovascular Magnetic Resonance
author_facet Lai Wyman W
Printz Beth F
Holland Brian J
author_sort Lai Wyman W
title Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
title_short Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
title_full Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
title_fullStr Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
title_full_unstemmed Baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
title_sort baseline correction of phase-contrast images in congenital cardiovascular magnetic resonance
publisher BMC
series Journal of Cardiovascular Magnetic Resonance
issn 1097-6647
1532-429X
publishDate 2010-03-01
description <p>Abstract</p> <p>Background</p> <p>One potential source of error in phase contrast (PC) congenital CMR flow measurements is caused by phase offsets due to local non-compensated eddy currents. Phantom correction of these phase offset errors has been shown to result in more accurate measurements of blood flow in adults with structurally normal hearts. We report the effect of phantom correction on PC flow measurements at a clinical congenital CMR program.</p> <p>Results</p> <p>Flow was measured in the ascending aorta, main pulmonary artery, and right and left pulmonary arteries as clinically indicated, and additional values such as Qp/Qs were derived from these measurements. Phantom correction in our study population of 149 patients resulted in clinically significant changes in 13% to 48% of these phase-contrast measurements in patients with known or suspected heart disease. Overall, 640 measurements or calculated values were analyzed, and clinically significant changes were found in 31%. Larger vessels were associated with greater phase offset errors, with 22% of the changes in PC flow measurements attributed to the size of the vessel measured. In patients with structurally normal hearts, the pulmonary-to-systemic flow ratio after phantom correction was closer to 1.0 than before phantom correction. There was no significant difference in the effect of phantom correction for patients with tetralogy of Fallot as compared to the group as a whole.</p> <p>Conclusions</p> <p>Phantom correction often resulted in clinically significant changes in PC blood flow measurements in patients with known or suspected congenital heart disease. In laboratories performing clinical CMR with suspected phase offset errors of significance, the routine use of phantom correction for PC flow measurements should be considered.</p>
url http://www.jcmr-online.com/content/12/1/11
work_keys_str_mv AT laiwymanw baselinecorrectionofphasecontrastimagesincongenitalcardiovascularmagneticresonance
AT printzbethf baselinecorrectionofphasecontrastimagesincongenitalcardiovascularmagneticresonance
AT hollandbrianj baselinecorrectionofphasecontrastimagesincongenitalcardiovascularmagneticresonance
_version_ 1725832814887174144