Humidity Effects on Fluidic Temperature Devices

A fluidic temperature device, designed by and obtained from The Harry Diamond Laboratories, was tested for stability when subjected to a variation of moisture content in an air stream. The device proved highly stable over a wide range of moisture contents. Slight variations in frequency were consist...

Full description

Bibliographic Details
Main Authors: John O. Geremia, Chih Wu
Format: Article
Language:English
Published: SAGE Publishing 1974-12-01
Series:Measurement + Control
Online Access:https://doi.org/10.1177/002029407400701204
id doaj-071a1f076093466ca1cafba8b27cb233
record_format Article
spelling doaj-071a1f076093466ca1cafba8b27cb2332020-11-25T03:17:37ZengSAGE PublishingMeasurement + Control0020-29401974-12-01710.1177/002029407400701204Humidity Effects on Fluidic Temperature DevicesJohn O. Geremia0Chih Wu1 Associate Professors, Mechanical Engineering Department, United States Naval Academy, Annapolis, Maryland 21402, USA Associate Professors, Mechanical Engineering Department, United States Naval Academy, Annapolis, Maryland 21402, USAA fluidic temperature device, designed by and obtained from The Harry Diamond Laboratories, was tested for stability when subjected to a variation of moisture content in an air stream. The device proved highly stable over a wide range of moisture contents. Slight variations in frequency were consistent with the variation predicted by the ideal gas behaviour of an air-water vapour mixture. The device is stable enough to be calibrated with dry air and corrected to any other humidity condition by application of the thermodynamic equations governing ideal gas mixtures.https://doi.org/10.1177/002029407400701204
collection DOAJ
language English
format Article
sources DOAJ
author John O. Geremia
Chih Wu
spellingShingle John O. Geremia
Chih Wu
Humidity Effects on Fluidic Temperature Devices
Measurement + Control
author_facet John O. Geremia
Chih Wu
author_sort John O. Geremia
title Humidity Effects on Fluidic Temperature Devices
title_short Humidity Effects on Fluidic Temperature Devices
title_full Humidity Effects on Fluidic Temperature Devices
title_fullStr Humidity Effects on Fluidic Temperature Devices
title_full_unstemmed Humidity Effects on Fluidic Temperature Devices
title_sort humidity effects on fluidic temperature devices
publisher SAGE Publishing
series Measurement + Control
issn 0020-2940
publishDate 1974-12-01
description A fluidic temperature device, designed by and obtained from The Harry Diamond Laboratories, was tested for stability when subjected to a variation of moisture content in an air stream. The device proved highly stable over a wide range of moisture contents. Slight variations in frequency were consistent with the variation predicted by the ideal gas behaviour of an air-water vapour mixture. The device is stable enough to be calibrated with dry air and corrected to any other humidity condition by application of the thermodynamic equations governing ideal gas mixtures.
url https://doi.org/10.1177/002029407400701204
work_keys_str_mv AT johnogeremia humidityeffectsonfluidictemperaturedevices
AT chihwu humidityeffectsonfluidictemperaturedevices
_version_ 1724630965762916352