Identification of mammography image artefact: A random assortment of low density blemishes

Following identification of an image artefact on a newly commissioned mammography system, the grid was initially not suspected as the cause as grid artefacts are generally associated with lines.  A reference phantom image was taken at an accredited centre was used to resolve the problem.  It is ther...

Full description

Bibliographic Details
Main Authors: A. Sweetlove, C. P. Herbst, C. Markgraaff, I. Nel
Format: Article
Language:English
Published: AOSIS 1997-03-01
Series:South African Journal of Radiology
Online Access:https://sajr.org.za/index.php/sajr/article/view/1642
Description
Summary:Following identification of an image artefact on a newly commissioned mammography system, the grid was initially not suspected as the cause as grid artefacts are generally associated with lines.  A reference phantom image was taken at an accredited centre was used to resolve the problem.  It is therefore felt to be essential for instructions not subject to peer evaluation to use phantoms.